11.2 Measuring mat
BIA tab not active
No seca patient file prepared
Prepare seca patient file
Measuring mat not activated
Administrator:
Activate measuring mat:
in the administrator
menu
No measuring mat present
Administrator, if desired:
• Retrofit measuring mat
• Activate measuring mat:
menu
Patient and user data can
not be restored manually
USB stick contains several backup
copies
Ensure that there is only one folder on the
USB stick with the designation
<ddMMyyyy_hhmm>_seca_monitor_b
ackup
(if necessary, rename other folders)
Backup copy on the USB stick has
been renamed
Ensure that backup copies are saved in the
folder with the designation
<ddMMyyyy_hhmm>_seca_monitor_b
ackup
(if necessary, rename folders)
Patient and user data
cannot initially be
encrypted after software
update
There is not enough memory on the
monitor for initial encryption
Initial manual encryption by administrator:
Exporting patient and user data
Restoring patient and user data
• If necessary, make previous device
settings (for example, WiFi connection)
manually
Fault
Cause
Remedy
Fault
Cause
Remedy
Measuring mat cannot be
switched on
Rechargeable battery discharged
Suspend measuring mat in storage
compartment of monitor and charge
rechargeable battery
Measuring mat faulty
Replace measuring mat
Charge status LED is red
Rechargeable battery discharged
Suspend measuring mat in storage
compartment of monitor and charge
rechargeable battery
Rechargeable battery faulty
Replace measuring mat
Inductive charging interface of monitor
covered, e.g. by other measuring
accessories
First suspend the measuring mat in the
magnetic catch of the monitor, then stow
other measuring accessories
Inductive charging interface faulty
Inform seca Service
WiFi LED is red
No WiFi connection to monitor
Administrator:
Check settings for WiFi connection and
correct if necessary
WiFi module of measuring mat faulty
Replace measuring mat
One or more Electrode
contact LEDs are red
Electrode cables not connected to
electrodes
Ensure that all electrode cables are
engaged in the push-buttons of the
electrodes
Electrodes defective
Replace electrodes
Electrode cables or measuring mat
faulty
Replace measuring mat