CON Brand

AG90 Propeller Replacement

Quick Start Guide

253.872.0284 

Page 1 of 4 

seametrics.com 

Warnings

• 

Confirm that U-clip retainer is installed and never remove the 

U-clip retainer when pipe is under pressure—may result in 

serious injury.

• 

Saddle bolts must be tightened evenly. Do not over tighten.

• 

Install security seals during installation if regulations require. 

• 

Improper sealing of cables or cable glands will void warranty.

• 

Ensure proper grounding when required.

• 

Programming pipe ID, pipe insertion hole size and straight pipe 

configuration is required for the meter to read. 

•   

To prevent tampering or changes to the program, after initial 

set-up, either enter a security pass-code, or remove the display 

lid and place a security tab over one of the arrows before 

installing the housing security seal.

Refer to instruction manual for further details.

Recommended Tools

Recommended:

• 

3/32” or 2.5mm (small) flat head screw driver

• 

Crescent Wrench

• 

5/32” Hex Key

Optional:

• 

Channel locks

• 

6” Strap Wrench

Check Components

Optional

Display Security Seal

Cable Gland

Security Clip & Seal

(Security Clip has small hole for seal wire)

Meter with Indicator 

and hinged cover

Saddle Assembly

(Packaged Separately)

Gasket

Power/Output Cable

Security Tab

Summary of Contents for AG90

Page 1: ...sertion hole size and straight pipe configuration is required for the meter to read To prevent tampering or changes to the program after initial set up either enter a security pass code or remove the display lid and place a security tab over one of the arrows before installing the housing security seal Refer to instruction manual for further details Recommended Tools Recommended 3 32 or 2 5mm smal...

Page 2: ...uts and tighten as shown below Torque to 75 ft lb in cross pattern BEFORE INSTALLING measure record inside diameter ID of pipe 10X 5X 2X Flow Flow Flow Flow Flow Flow Upstream straight pipe is selected during initial setup Upstream options are 5X or 10X the diameter and are based on the amountofstraightpipeavailableineitherneworpropellermeter replacement installation Downstream straight pipe requi...

Page 3: ...que cable gland sealing nut to 22 in lbs Clamp cable with strain relief clips Attach drain wire lug to bracket post NOTE If not using external output no wiring is required With No External Output White C1 Pulse Green C1 Pulse C1 With Pulse Output Only Pulse with External Power White C1 Pulse Green C1 Pulse C1 Red C1 DC Black C1 DC Unscrew the display lid and remove it from the meter Unsnap the dis...

Page 4: ...h a 5 second sampling rate is four years Increase sampling interval to extend battery life May vary depending on environmental factors See instruction manual for more details EXIT To SUBMENU The EXIT tab in the MAIN MENU has a second function Tap five times to enter a SUBMENU screen from which you can access the following additional options T UNIT R UNIT SET P DAMP SET 4 SET 20 SET F EXIT PRESS EX...