To move the data in any direction, push the "Ctrl" key on the PC keyboard and
the left button of the mouse pointer at the same time.
•
To select a specific part of the data to zoom in on, pull the mouse pointer
diagonally.
•
Push
Auto Range
to see the data for each selected parameter. The software
adjusts the scale so that the data will always show.
•
Push
Default Ranges
to go back to the manufacturer-set default scale for each
parameter.
•
Put a check in the box next to "Show Data Points" to see the value of the
collected data when the mouse moves over each point.
•
Push
Select Sensors To Display
to change the parameters to look at in the
Time Series
graph.
16.
Push
Stop
. The "Connection Mode" shows "Setup."
4.1 Batteries
4.1.1 Alkaline batteries
Twelve D-cell alkaline batteries are installed in the SeaFET by the manufacturer. The
user should verify that the batteries are installed correctly and supply the specified
voltage. Refer to
Remove and replace alkaline batteries
on page 32 for details about
how to change the batteries when necessary.
4.1.2 Lithium batteries
N O T I C E
Use only the batteries recommended by the manufacturer as replacements. Do not mix
chemistries, V-Ah, or new and used batteries.
The manufacturer ships the twelve lithium batteries for the SBE-37-SMP-ODO separately.
Do the steps below to install or replace the batteries. See also the "how to" video on the
manufacturer's website to do this procedure.
Table 2 Recommended lithium battery brands
SAFT LS-14500 (included)
3.6 V, 2.6 Ah
Tadiran TL-4903
3.6 V, 2.4 Ah
Electrochem BCX85 series
3.9 V, 2.0 Ah
1.
Make sure that the end flange and pressure housing are dry.
2.
Use a 9/64" hex key to remove the two screws on the sides of the pressure housing.
3.
Install these two screws into the sides of the end flange to start to loosen the end
flange.
4.
Turn the end flange counter-clockwise to loosen it from the pressure housing.
5.
Pull gently to disconnect the battery wires in the end flange from the battery pack.
6.
Use a lint-free tissue to remove any water from the O-ring surfaces inside the
pressure housing and end flange.
Set up sensor and verify function
16
Summary of Contents for Deep SeapHOx V2
Page 2: ......
Page 8: ...2 1 3 Illustrations Figure 1 SeaFET V2 dimensions Specifications 6 ...
Page 9: ...Figure 2 Y cable for SeapHOx V2 Figure 3 Data I O cable Specifications 7 ...
Page 12: ...Specifications 10 ...
Page 16: ...Product overview 14 ...
Page 26: ...Deployment and recovery 24 ...
Page 42: ...Maintenance 40 ...
Page 46: ...Reference 44 ...
Page 50: ...Software reference 48 ...
Page 55: ......