Section 4: Operating System
SBE 45 Interface Box
21
5.
In the Start Real-Time Data Acquisition dialog box, click
Start
.
A.
If you selected
Begin archiving data immediately
or
Begin archiving
data when ‘Start Archiving’ command is sent
above, and selected
Prompt for Header Information
in the Header Form setup (Configure
Outputs), the Header Information dialog box appears. Fill in the
desired header and click OK.
B.
If you selected
Check Scan Length
in the Options menu, Seasave
checks the configuration (.xmlcon or .con) file to verify that the scan
length defined by the configuration file matches the SBE 45 (i.e.,
number of outputs and inclusion of NMEA is as defined in the
configuration file). If a
Scan length error
appears, verify that:
•
You are using the correct .xmlcon or .con file.
•
The .xmlcon or .con file has been updated as necessary if you
added or deleted the SBE 38, added or deleted NMEA, etc.
C.
Seasave sends a message:
Waiting for data . . .
. Seasave will
time out
if
data is not received within
Timeout in seconds at startup
.
D.
Real-time data then starts appearing in the screen display(s).
6.
To stop real-time data acquisition: In the Real-Time Data menu,
select
Stop
.
7.
Stop the SBE 45 sampling by turning off power to the Interface Box.
Processing Data with SBE Data Processing
1.
Convert the .hex data file created by Seasave to a .cnv file in SBE Data
Processing’s Data Conversion module.
2.
Once the data is converted, it can be further processed and plotted in
SBE Data Processing’s other modules. Note that if the data stream
includes SBE 38 remote temperature data, SBE Data Processing’s Data
Conversion and Derive modules use the remote temperature data when
calculating density and sound velocity.
Note:
See the SBE Data Processing
manual and/or Help files.
83
Summary of Contents for SBE 45 MicroTSG
Page 2: ...This page intentionally left blank ...
Page 4: ...Manual Generation Date 13 February 2012 1 ...
Page 6: ...This page intentionally left blank 3 ...
Page 7: ...4 ...
Page 112: ...109 ...
Page 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Page 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Page 129: ...126 ...
Page 130: ...127 ...
Page 131: ...128 ...
Page 132: ...129 ...
Page 133: ...130 ...
Page 134: ...131 ...
Page 135: ...132 ...
Page 136: ...133 ...
Page 137: ...134 ...
Page 138: ...135 ...