Section 3: Setting Up System
SBE 45 Interface Box
14
Commands
This section describes commands and provides sample outputs.
When entering commands:
•
Input commands to the Interface Box in upper or lower case letters and
register commands by pressing the Enter key.
•
The Interface Box sends
?
if an invalid command is entered.
•
If the system does not return an
S>
prompt after executing a command,
press the Enter key to get the
S>
prompt.
•
The Interface Box can accept commands while transmitting data. Type in
any command and press the Enter key; the
S>
is not required.
Status
Command
DS
Display firmware version and baud rate
settings. Equivalent to Status button
on Toolbar.
List includes, where applicable, command
used to modify parameter.
•
firmware version
•
SBE 45 baud rate [
45Baud=
]
•
SBE 38 baud rate [
38Baud=
]
•
NMEA baud rate [
NMEABaud=
]
•
output format [
Format=
] –
standard output format
or
output format 1
Baud Rate
Commands
PCBaud=x x=
baud rate for communication between
Interface Box and computer
(4800, 9600, or 19200). Default 9600.
45Baud=x x=
baud rate for communication between
Interface Box and SBE 45 (1200, 2400,
4800, or 9600). Default 4800. Must match
baud set in SBE 45 with
Baud=
.
38Baud=x x=
baud rate for communication between
Interface Box and optional SBE 38 (1200,
2400, 4800, or 9600). Default 4800. Must
match baud set in SBE 38 with
Baud=
.
NMEABaud=x x=
baud rate for communication between
Interface Box and optional NMEA
navigation device (4800 or 9600).
Default 4800. Must match baud of
NMEA navigation device.
Example:
(user input shown in bold)
S>
DS
SBE45 Junction Box V 1.4
SBE 45 baud rate = 4800
[
45Baud=
]
SBE 38 baud rate = 4800
[
38Baud=
]
NMEA baud rate = 4800
[
NMEABaud=
]
standard output format
[
Format=
]
Note:
Baud rate for communication
between Interface Box and
computer must be greater than baud
between Interface Box and SBE 45,
SBE 38, and NMEA navigation
device.
76
Summary of Contents for SBE 45 MicroTSG
Page 2: ...This page intentionally left blank ...
Page 4: ...Manual Generation Date 13 February 2012 1 ...
Page 6: ...This page intentionally left blank 3 ...
Page 7: ...4 ...
Page 112: ...109 ...
Page 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Page 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Page 129: ...126 ...
Page 130: ...127 ...
Page 131: ...128 ...
Page 132: ...129 ...
Page 133: ...130 ...
Page 134: ...131 ...
Page 135: ...132 ...
Page 136: ...133 ...
Page 137: ...134 ...
Page 138: ...135 ...