SCS TMTOOL Installation, Operation And Maintenance Download Page 1

TB-9048

 Page 1 of 3

© 2016 DESCO INDUSTRIES, INC.

Employee Owned

SCS 

- 926 JR Industrial Drive, Sanford, NC 27332 • (919) 718-0000 • Website: 

StaticControl.com

Verification Tester for WSMONITOR

Installation, Operation and Maintenance

January 2016

 USER GUIDE TB-9048

Description

The SCS TMTOOL Verification Tester is designed to 

simplify the process of calibrating SCS WSMONITOR 

series Workstation Constant Monitors.  The SCS 

TMTOOL Verification Tester provides an efficient and 

accurate means to perform ANSI/ESD S20.20 required 

Compliance Verification per ESD TR53.

Frequency of recalibration should be based on the 

critical nature of those ESD sensitive items handled and 

the risk of failure for the ESD protective equipment and 

materials.  In general, SCS recommends that calibration 

be performed annually.

Use the SCS TMTOOL Verification Tester to perform 

compliance verification testing of the SCS WSMONITOR 

series.  It can be used at the workstation where the 

monitors are installed, virtually eliminating downtime. 

Testing takes a few minutes, verifying that the monitor is 

operating within tolerances (adjustment is not possible).  
The TMTOOL Verification Tester is to be used with the 

following items:

WSMONITOR1

WSMONITOR1-230VAC

WSMONITOR2

WSMONITOR2-230VAC

WSMONITOR3

WSMONITOR3-230VAC

Packaging

1  Verification Tester

1  Certificate of Calibration

Operation

WSMONITOR1

1. 

Ground the TMTOOL Verification Tester by 

connecting its GROUND lead to electrical ground. 

2. 

Power the WSMONITOR1 monitor.

 

3.  Insert the OPERATOR LEFT lead from the 

Verification Tester into the OPER jack located on 

the front of the WSMONITOR1 monitor.

 

Made in the

United States of America

Figure 1. SCS 

TMTOOL

 Verification Tester

Figure 2. Connecting the Verification Tester to the 

WSMONITOR1 monitor

4. 

Press and hold the OPERATOR LEFT button on the 

Verification Tester.  The LED on the WSMONITOR1 

should illuminate green.  

Release the button.  The 

LED should flash red, and the audible alarm should 

sound briefly.

 

5.  The monitor is within specifications if it responds as 

described above.

WSMONITOR2

1. 

Ground the TMTOOL Verification Tester by 

connecting its GROUND lead to eletrical ground. 

2. 

Power the WSMONITOR2 monitor.

 

3.  Insert the OPERATOR LEFT lead from the 

Verification Tester into the operator jack located on 

the front of the WSMONITOR2 monitor.

 

4.  Insert the BENCH lead from the Verification Tester 

into the BENCH SYSTEM GROUND jack located on 

the back of the WSMONITOR2 monitor.

 

Reviews: