background image

TB-9083

 Page 1 of 3

© 2020 DESCO INDUSTRIES INC

Employee Owned

SCS 

- 926 JR Industrial Drive, Sanford, NC 27332

East: (919) 718-0000 | West: (909) 627-9634 • Website: 

StaticControl.com

June 2020

Figure 1. SCS 

770065

 Verification Tester

Description

The SCS 

770065

 Verification Tester is used to 

perform periodic test limit verification of the SCS 

724

 

Workstation Monitor, 

724 Plus

 Workstation Monitor, 

and 

725

 Portable Wrist Strap Monitor. Verification 

may be accomplished without removing the monitor 

from its workstation. The Verification Tester is 

National Institute of Standards and Technology (

NIST

traceable. Frequency of verification is based on the 

critical nature of the ESD susceptible items handled. 

SCS recommends annual calibration of workstation 

monitors and the Verification Tester. The SCS 

770065

 

Verification Tester meets ANSI/ESD S20.20 and 

Compliance Verification ESD TR53.
The SCS 

770065

 Verification Tester can be used with 

the following items:

Item

Description

724

Workstation Monitor

724MO

Workstation Monitor, No Power Adapter

725

Portable Wrist Strap Monitor

770724

724 Plus Workstation Monitor

Packaging

1  Verification Tester

1  Certificate of Calibration

Operation

724 Workstation Monitor

Testing the Operator Circuits

The operator high test limit and test voltage may be 

configured using the switches located on the right-side 

of the 724 Workstation Monitor. The test limit may be 

set to either 10 megohms or 35 megohms, and the test 

voltage may be set to either 9V or 16V. Take note of the 

test limit setting before following the instructions listed 

below. The test voltage may be set to any position.

Verification Tester

Operation Instructions

Made in the

United States of America

 USER GUIDE TB-9083

Figure 2. Operator test limit and test voltage switches 

located on the right-side of the 724 Workstation Monitor

The Verification Tester’s ground clip toggle switch is not 

used in the verification of the 724 Workstation Monitor. It 

may be set to any position.
1.  Insert the Verification Tester’s stereo plug into the 

monitor’s operator remote jack #1. 

Figure 3. Using the Verification Tester with the 724 

Workstation Monitor