Description
12
GA01423_002_C0 - 10/2016 - © Leybold
1.3.1 Motor Data
Motor data
SC 5 D
Type
Single-phase induction motor / Motor 4P / totally enclosed
insulation class B IP 44
thermal protector TP 212 - automatic reset type
Power consumption
W
150
Voltage
V 100 115
200
230
Rated current
50 Hz
A
2.6
–
1.3
1.6
60 Hz
A
2.1
2.2
1.1
1.1
Speed at
50 Hz
min
-1
1430 –
1430 1440
60 Hz
min
-1
1730 1740
1730
1740
Noise level
2
at 1m dB(A)
≤ 52
with purge gas open (ON)
dB(A)
≤ 57
Direction of inlet
Vertical
Dimensions (L x W x H)
mm
308 x 214 x 225
Weight
kg 14
Part numbers
133 100 133 000
Motor Data
SC 15 D
Type
Single ph.
induction motor 4P
Three-phase
induction motor
5
4P
totally enclosed
totally enclosed
insulation class B / IP 44
insulation class B / IP 44
thermal protector TP 212
automatic reset type
Power consumption
kW
0.4
0.4
Voltage
V
100 115 200 230
200 208 230 380 400 415 460
Rated current
50 Hz
A
4.8
-
2.6
2.4
1.6
-
-
0.9
0.9
1.0
-
60 Hz
A
4.8
4.3
2.8
2.4
1.9
1.9
1.8
-
-
-
1.0
Speed at
50 Hz
min
-1
1440
-
1430
1450
1420
-
-
1440 1440 1440
-
60 Hz
min
-1
1710 1740 1700
1730
1660
1660 1690
-
-
-
1720
Noise level
2
at 1m dB(A)
≤ 58
with purge gas open (ON)
dB(A)
≤ 66
Direction of inlet
Vertical
Dimensions (L x W x H)
mm
400 x 252 x 336
370 x 252 x 336
Weight
kg
25
23
Part numbers
133 101 133 001 133 003