background image

TS-1078

Tamp Swing Arm

Replacement

(MP2/MP2+)

Document Owner:

Service Team

Effective Date:

AUG-24-2022

Page 1 of 5

Purpose

This document explains the process for replacing the swing arm on a Sci-Print MP2 or MP2+.

Intended Use

When the swing arm requires changing.

Required Material and Equipment

Sci-Print MP2 or MP2+

2mm allen wrench

3mm allen wrench

Intended Audience

Certified Scinomix Technician

Prerequisites

N/A

Procedure

1. Power off instrument

2. Turn air supply to instrument off

3. Using a 2mm allen wrench remove the four screws for the swing arm tower housing, and

set housing aside.

Note

: The fourth screw (not seen in Figure 1 below) is behind the swing arm; swing it to

the other side for access.

File Location: Google Drive

Revision: A   Approved Date: AUG-23-2022   Approved by: Service

Reviews: