10 

Transmission 

method: 

Asynchronous 

signal

transmission in start-stop-operation.

Mode  of  transmission:  Fully  Duplex.  1  start  bit;  7

character bits; 1 parity bit [straight (even)); 1 stop bit

Transmission speed: 9600 bit/s

Note: 

Forbid  to  insert  or  remove  the  RS232C 

communication line when switch on! 

10 Faults 

Instruments can’t be powered ON

-

 

Check whether the power line is unplugged 

-

 

Check whether the fuse is broken or loose 

Fault in power during self-test

-

 

Switch  OFF  the  unit,  then  switch  ON  and  reset 

the instruments to factory default setting. 

Stir speed cannot reach set point

-

 

Excessive  medium  viscosity  may cause abnormal 

speed reduction of the motor 

Unit cannot be powered off when switched off.

-

 

Check if the residual heat warning function is still 

ON and hotplate temperature is above 50

 (the 

LCD still work and “Hot” flash). 

If  these  faults  are  not  resolved,  please  contact 

manufacturer/supplier. 

11 Maintenance and Cleaning 

Correct  usage  and  maintenance  of  the  equipment  can 

extend  the  service  life  of  equipment.  Please  keep  the 

instrument  dry  and  clean,  quickly  remove  and  overflowing 

liquid,  use  non-abrasive  cleaner  to  clean  up  the  outer 

surface, please do not connect with the power supply if wet. 

If  wet  solids  or  liquid  enter  into  the  instrument,  please 

disconnect  the  power  supply  and  do  not  use,  contact  the 

manufacturer/supplier for more advice. 

Do  not  use  a  spray  cleanser  on  the  instrument  when

cleaning.

Unplug the power cord when cleaning.

Only use recommended cleansers:

Dyes 

Isopropyl alcohol 

Construction 

materials 

Water  containing  surfactant  / 

Isopropyl alcohol 

Cosmetics 

Water  containing  surfactant  / 

Summary of Contents for MS-H-Pro-T

Page 1: ...imer Please read the User Manual carefully before use and follow all operating and safety instructions Technical specifications and outline are subject to change without prior notice www scilogex com...

Page 2: ...ontrol 4 4 1 Control elements 4 4 2 Display 6 5 Trial Run 6 6 Function Heating 7 6 1 Working with the PT1000 external temperature sensor 8 6 2 Residual heat warning HOT 8 7 Stirring 9 8 Timer 9 9 Remo...

Page 3: ...ion Serial number on the rear panel Certification Description of problem i e hardware or software Methods and procedures adopted to resolve the problems Your contact information Warranty This instrume...

Page 4: ...reduce the speed if Stirring bar breaks away due to high speed The instrument is not running smoothly or the container is not centered on the plate Temperature must always be set to at least 50C lowe...

Page 5: ...be operated by adults Keep away from high magnetic fields 2 Proper Use The instrument is designed for mixing and or heating liquids in laboratories Do not use more than 2000 meters above sea level Te...

Page 6: ...parent damage to the unit do not attempt to operate 3 2 Listing of Items The package includes the following items Items Qty Main unit 1 Power cable 1 Stirrer bar 1 User Manual 1 Table 1 4 Control 4 1...

Page 7: ...f Parameter setting knob At the parameter setting interface set the parameter by turning the knob Pressing the knob turns on the initializer LCD display LCD displays the real time working state and al...

Page 8: ...sing remote control Rated value Real value Display values in when heating or stirring or timer function is switched ON Table 3 Note If both heating and stirring functions have been started at the same...

Page 9: ...rcuit The hotplate temperature can also be controlled from a separate adjustable safety circuit The two temperature sensors PT1000 internal for temperature control are built into the hotplate The sing...

Page 10: ...ature more precisely The external temperature sensor must be put in sample container if detect anomalies heating will automatically shut down Please do the following at this time Turn off the power En...

Page 11: ...he parameter set knob to target time press parameter set knob to switch on timer function Attention The heating and stirring function can be switched on at the same time only when setting the timer fu...

Page 12: ...the residual heat warning function is still ON and hotplate temperature is above 50 the LCD still work and Hot flash If these faults are not resolved please contact manufacturer supplier 11 Maintenan...

Page 13: ...following safety standards EN 61010 1 UL 3101 1 CAN CSA C22 2 1010 1 EN 61010 2 10 Construction in accordance with the following EMC standards EN 61326 1 Associated EU guidelines EMC guidelines 89 33...

Page 14: ...l with ceramic coated Dimensions of workplate mm 135 Heating power W 500 Temperature range C Room temperature 340 Step 1 Temperature display C LCD Temperature display accuracy C 0 1 The safety tempera...

Page 15: ...0V 240V 50Hz 60Hz Cat No Descriptions Accessories 18900016 PT1000 A Temperature sensor for digital hotplate model used for MS H Pro MS H ProT MS7 H550 Pro MS7 H550 S length of 230mm 18900136 PT1000 B...

Page 16: ...tion block for 500ml round bottom flask one flask capacity 18900006 Stirrer bar 10mm x 6mm 1pc pk 18900007 Stirrer bar 15mm x 8mm 1pc pk 18900008 Stirrer bar 20mm x 8mm 1pc pk 18900009 Stirrer bar 25m...

Page 17: ...SCILOGEX LLC 1275 Cromwell Ave Suite C6 Rocky Hill CT 06067 USA Tel 1 860 436 9221 Fax 1 860 436 9745 info scilogex com www scilogex com...

Reviews: