4.7 

Running a Test 

 

1.  Obtain a serum or lithium heparinized plasma 

sample. 

2.  On the main screen, touch Worklist or Manual 

Test. Figure 3 

3.  In Worklist Mode, confirm that all fields have 

correct information and then touch   to proceed. 
In Manual Mode, enter sample information in the 
required fields. Touch   to proceed. Figure 4 

 

Figure 3 

 

Figure 4 

 

4.  Open the pouch by tearing at the notch. 

Carefully remove the cartridge by the handle and 
place it on a flat surface. Figures 5 & 6 

 

NOTE:  

Do not touch the bottom of the cartridge. 
 

NOTE:  

If the cartridge was refrigerated, allow to warm to 
room temperature for at least 15 minutes before 
opening the pouch. 
 

NOTE:  

Cartridge must be used within 1 hour of removal from 
pouch. 
 

 

Figure 5 

 

Figure 6 

5.  With the cartridge flat on the bench, remove the 

pull-tab from the sample inlet port and discard. 
Use the cartridge handle to steady the flat 
cartridge while removing the pull-tab. Figure 7 

 
 
 
 

NOTE:  

Cartridge must be used within 15 minutes of 

removing pull-tab. 

 

 

6. 

NOTE INSTRUCTIONS FOR T4 ONLY (OTHER TESTS SKIP TO 7.) 
Dilute patient sample using supplied Heska T4 Diluent. 
a. 

Use supplied 100 µl fixed volume mini-pipette to aspirate 100 µl of serum or 
plasma sample. 

b. 

Flick or tap the Heska T4 Diluent tube to ensure there is no liquid adhering to the 
Diluent tube inner cap surface.

 

c. 

Dispense the full 100 µl patient sample into the Diluent tube. Ensure proper 
mixing by then aspirating and dispensing 5x using the same pipette tip. 

d. 

Aspirate 100 µl of the diluted sample using the mini-pipette and same pipette tip, 
insert the pipette tip into the T4 cartridge sample inlet port hole, and dispense the 
full amount into the hole. (Skip to 8.) 

Summary of Contents for Element i+

Page 1: ...User Manual Immunodiagnostic Analyzer Element i For Veterinary In Vitro Diagnostic Use Only...

Page 2: ...t prior notice Heska MBio shall not be liable for malfunctions and damage resulting from installation relocation remodeling maintenance and repair done by other than dealers specified by Heska Heska M...

Page 3: ...8 2 3 Calibration 8 3 Transport Unpacking and Installation 9 3 1 Packaging 9 3 2 Unpacking 9 3 3 Parts List 9 4 Analyzer Operation 10 4 1 Turning the Analyzer On and Off 10 4 2 Main Screens and Contro...

Page 4: ...use only The intended use is to quantify biological markers in animal serum or plasma samples using Element i Cartridges Do not use the equipment for other purposes Please refer to instructions includ...

Page 5: ...s with Part 15 of FCC regulations and FDA regulations 21 CFR 1040 10 and 1040 11 except for deviations pursuant to Laser Notice No 50 NOTE This equipment has been tested and found to comply with the l...

Page 6: ...or be opened when a cartridge is not present If the enclosure is damaged immediately stop using the analyzer to prevent possible exposure to hazardous laser radiation CAUTION Use of controls or adjust...

Page 7: ...1 Components and Labeling Front Analyzer label found on the bottom of analyzer Back Touch screen Status light Cartridge port door Carrying handle Air filter cover and filter Power button USB A2 0 port...

Page 8: ...n Laser Class 1 per IEC 60825 1 Operating temperature 59 86 F 15 30 C indoors Operating relative humidity 25 65 non condensing Storage temperature 32 113 F 0 45 C Storage humidity 5 95 non condensing...

Page 9: ...ement i Analyzer and place on a level stable surface Inspect for any obvious signs of damage Report any damage immediately Remove protective film from touchscreen if present Before using allow the ana...

Page 10: ...hen initially powered the analyzer will run through a series of self diagnostic checks When boot up is complete the indicator light will glow steady and the login screen will appear The analyzer shoul...

Page 11: ...s you need them If you want to set the reference areas you must first go to the settings Then go to the administrator and there you will find the button reference groups In this menu you can change th...

Page 12: ...until this file is loaded Use the following procedure to load the CDF 1 Insert the provided USB flash drive Figure 1 with cartridge lot calibration information into any of the four USB ports on the b...

Page 13: ...ingerprints smudges and other contamination Fluorescence illumination and imaging are through the clear plastic component on the bottom of the cartridge 4 6 Running a Sample This section assumes that...

Page 14: ...e 6 5 With the cartridge flat on the bench remove the pull tab from the sample inlet port and discard Use the cartridge handle to steady the flat cartridge while removing the pull tab Figure 7 NOTE Ca...

Page 15: ...re Sample screen to open the analyzer door Figure 9 Insert the cartridge until you feel a click and hear a beep Figures 10 11 Figure 9 Figure 10 Figure 11 9 The test will run automatically A status ba...

Page 16: ...n Figure 13 The screen will provide an indication when it is safe to remove the used cartridge Figure 14 4 8 Review Test Results Figure 13 Figure 14 1 Click on the button Review Tests 2 Now you can se...

Page 17: ...the analyzer can be cleaned by wiping with a mild detergent 70 solution of ethanol or isopropyl alcohol in water or 10 bleach solution CAUTION Do not directly spray the Analyzer with cleaner Apply cle...

Page 18: ...new cartridge Unable to read the barcode from the inserted cartridge Barcode not read Verify that barcode is not damaged and retry the cartridge If error persists select a new cartridge The inserted c...

Page 19: ...ed error editing reference range General error Confirm reference ranges are set in the Administrator settings Unexpected error deleting reference range General error Confirm reference ranges are set i...

Reviews: