background image

 

 

 

 

 

 

 

 

Test Tube pH Pen 
  

SP-850062, SP-850063, 

SP-850065, SP-850066

,  

SP-850067

  

Summary of Contents for SP-850062

Page 1: ...Test Tube pH Pen SP 850062 SP 850063 SP 850065 SP 850066 SP 850067...

Page 2: ...Test Tube pH Pens Models SP 850062 63 65 66 67 are designed for easy calibration and simple accurate measurement The 12 cm long probe is capable of testing samples even in narrow mouthed containers Th...

Page 3: ...5 units of 4 7 and 10 Measuring stability indicator Protective storage bottle with KCI storage solution for sensor Low battery indicator Automatic shutoff Dual pH and temperature display Models SP 85...

Page 4: ...Y SP 850062 1 pH reading 2 Data hold indicator 1 3 Calibration 4 Low battery icon SP 850063 65 66 67 1 pH reading 2 Data hold indicator 3 Calibration 1 4 Automatic Temperature Compensation ATC 5 Low b...

Page 5: ...to 60 mV pH 10 00 250 mV to 97 mV Calibration Procedure 1 Press POWER button to turn the meter on 2 Pour the pH buffer into a clean container and place the electrode into the buffer We recommend star...

Page 6: ...d 10 buffers CUSTOM CALIBRATION During calibration mode you can adjust the buffer values to non standard values For example 4 00 pH may be adjusted from 3 50 to 4 50 7 00 pH may be adjusted from 6 50...

Page 7: ...splay on the screen 5 End displays and the meter returns to Normal Mode 6 Rinse the probe with deionized water a rinse solution or tap water 7 Repeat steps 2 7 to calibrate the meter with remaining bu...

Page 8: ...ng the probe to dry out It is normal for white crystals to be present on the cap or electrode assembly 1 Unscrew the storage bottle and remove it from the probe exposing the electrode 2 Place the elec...

Page 9: ...urement Mode 6 The pH value displays on the LCD Models SP 850063 65 66 67 also display the temperature 7 Press POWER button to turn the meter off 8 Rinse the electrode with deionized water a rinse sol...

Page 10: ...Data Hold Function 1 During measurement press HOLD to freeze the reading on the display H appears at the top of the LCD and the pH decimal point stops flashing 2 Press HOLD to return to Normal Mode 10...

Page 11: ...n 1 If the meter is on press POWER button to turn the meter off 2 Press both POWER button and HOLD for two seconds 3 The meter will turn on and n will display 4 The meter returns to Normal Mode Automa...

Page 12: ...either C or F To select the temperature unit 1 If the meter is on press POWER to turn the meter off 2 Press and hold both POWER and CAL until the temperature unit appears on the LCD 3 Press HOLD to t...

Page 13: ...are correctly inserted Remove the batteries for one minute and then reinstall them Replace the old batteries with new ones Clean the probe by immersing the electrode in tap water for 10 15 minutes Ri...

Page 14: ...g displays of range the temperature is H or L too high or too low Model SP 850063 BATTERY INSTALLATION AND REPLACEMENT Note Remove the batteries when the meter will not be used for one month or longer...

Page 15: ...e back of the meter by using the slot to turn it clockwise The cover will now come off easily when you turn the timer over Take care not to discard the washer in the battery compartment 3 Remove the o...

Page 16: ...Replace the battery cover positioning the tabs so that the cover is flush with the meter 5 Secure the cover by turning it counter clockwise Note Recalibrate the meter after replacing the batteries 16...

Page 17: ...of the electrode The pH tip must always be kept moist Place it in the protective storage bottle containing KCI storage solution whenever it is not in use Rinse the electrode in deionized water a rinse...

Page 18: ...4 00 0 01 0 1 2 12 00 0 01 0 1 Temperature Operating Temperature Operating RH Storage Temperature Storage RH LCD Display Dimensions Weight 0 50 C 32 122 F 0 1 C F 1 C 2 F 0 50 C 32 122 F 80 10 50 C 14...

Page 19: ...esulting from accident misuse or abuse of the product To obtain warranty service contact your Distributor SCIGIENE CORPORATION Tel 416 261 4365 Email support scigiene com Be sure to include a descript...

Page 20: ...Rev 03 01 2019 20...

Reviews: