Figure 3-5 Diverter Valve—Diverter Mode Position B
Description
Item
To mass spectrometer
1
From column
2
Waste out
3
Calibrant Delivery System
The calibrant delivery system (CDS) introduces calibration solution for automated mass calibration of the mass
spectrometer, to ensure that the mass accuracy of the system is maintained throughout batch acquisition.
Because calibration takes about a minute and a half, we recommend frequent calibration.
Replace the CDS Bottle
WARNING! Toxic Chemical Hazard. Refer to the chemical product Safety Data Sheets
and follow all of the recommended safety procedures when handling, storing, and
disposing of chemicals. For health and safety precautions, refer to
.
The CDS supports up to two bottles of calibrant. Use bottle one for the positive calibrant solution. Use bottle two
for the negative calibrant solution. Be sure to install the bottle in the correct position to avoid cross-contamination.
SCIEX X500 QTOF System
System User Guide
41 / 234
RUO-IDV-05-2334-E
Operating Instructions—Hardware