Change Sample Racks
The system is provided with 1.5 mL sample vial plates as a standard. This section provides the procedure for
changing the settings to use other plates.
Set the Rack ID
The rack ID is the ID the module uses to identify the type of plate that has been inserted. Set the rack ID by operating
the sliding switch at the rear of the sample rack.
The two sample racks provided with the module have rack IDs set to A and B by default, which means that the
needle strokes and rack positions are adjusted for a 1.5 mL sample vial plate.
Note:
If the rack ID settings have been changed, then it is necessary to configure the plate type, set the needle
stroke from the top, and then adjust the rack position. Refer to the
ExionLC
TM
100 System Hardware User
Guide
for instructions.
Note:
If the plate type is changed but the same rack ID is used, then the needle stroke from the top should be
set and the rack position adjusted.
1. Remove the sample rack (A).
Figure 5-23 Sample Rack Removal
Hardware User Guide
ExionLC
™
100 System
RUO-IDV-05-1839-C
124 / 219
Service and Maintenance