8 LOW CURRENT DECK
The Low Current Deck Option for the 5000 Series Tester, with matching software,
extends the low-end leakage current measurement range to 100
ρA with 50ρA
resolution. The lower current ranges are attainable by virtue of the measurement
circuitry of the Low Current Deck. This places the measurement circuit as close as
possible to the device being measured, thereby minimizing both leakage and AC
pick-up.
The Low Current Deck also provides programmable soak time to 98 milliseconds,
Kelvin lead continuity testing and integration of the test answer. The “integrate”
mode measures the parameter over a full AC line cycle to reduce the effects of AC
pick-up.
The device under test is plugged into the Low Current Deck panel for manual testing.
A cable to attach the test leads to a handler is optionally available.
The Low Current Deck comes in two cases; the stand-alone case with 50-Pin data
cable and device cable and plug-in case with 50-Pin data cable.
8.1 Installation
Connection of the stand-alone case Low Current Deck is as shown in Figure 8-1.
The plug-in case connects directly to the front panel of the 5000 Series Tester. In
both case, the 50-Pin connector connects from the back panel of 5000 Series Tester
and the Low Current Deck.
NOTE: Substantial force is required to plug the Low Current Deck directly into the
panel of the 5000 Series tester. This assures good connection.
Use the supplied shielded cable to connect the front panel of the 5000 Series Tester
to the back of the Low Current Deck. The length of this cable is significant in the
measurement of low currents.
If the Low Current Deck was not part of your original configuration, or if you want to
reconfigure your system, you must set switch #4 (dipswitch located on the Analog
Board of the 5000 Series Tester). Position switch #4 ON to enable or OFF to disable
the Low Current Deck. Any changes made to the switch setting should also be made
to the “Mainframe Configure” of the PCW software (see Section 6 for more
information).
If connecting to a handler, use a cable set with a maximum length of six feet. This
cable is optionally supplied.
Cut the cable as short as possible
. Even though the
cable is shielded, with the shield grounded (on the tester end only), AC pick-up and
cable charging time is minimized. The most sensitive test leads for low current
For the system to operate properly the hardware
configuration of the system must match the
dipswitch settings on the Analog Board and the
configuration setting in the PCW software!
Summary of Contents for 5000 Series
Page 2: ......
Page 4: ......
Page 10: ......
Page 30: ...TABLE OF CONTENTS xx Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 38: ...1 OVERVIEW 1 8 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 52: ...3 THEORY 3 6 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 62: ...4 SPECIFICATIONS 4 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 64: ...5 BASIC OPERATION 5 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 102: ...6 PC SOFTWARE 6 38 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 218: ...7 MAINTENANCE TROUBLESHOOTING 7 116 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 226: ...8 LOW CURRENT DECK 8 8 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 234: ...10 PIN PROGRAMMABLE SCANNER 10 4 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 242: ...APPENDICES 13 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 244: ...APPENDIX A 13 4 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 250: ...APPENDIX C 13 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 252: ...APPENDIX D 13 12 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 254: ...APPENDIX E 13 14 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 258: ...APPENDIX F 13 18 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 262: ...APPENDIX G 13 22 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 268: ...APPENDIX H 13 28 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 290: ...APPENDIX I 13 50 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 292: ...APPENDIX J 13 52 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 296: ...APPENDIX K 13 56 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 304: ...APPENDIX M 13 64 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 308: ...APPENDIX N 13 68 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 312: ...APPENDIX P 13 72 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 316: ...14 APPLICATION NOTES 14 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 318: ...APP 001 14 4 Version 1 23 MODEL 5000 SERIES MANUAL Figure 14 2 APP 001 Test Screen ...
Page 320: ...APP 001 14 6 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 322: ...APP 002 14 8 Version 1 23 MODEL 5000 SERIES MANUAL Figure 14 6 APP 002 Bin Sort Screen ...
Page 324: ...APP 003 14 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 326: ...APP 004 14 12 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 332: ...APP 006 14 18 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 336: ...APP 007 14 22 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 338: ...APP 008 14 24 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 340: ...APP 009 14 26 Version 1 23 MODEL 5000 SERIES MANUAL ...