APPENDIX H
13-26
Version 1.23
MODEL 5000 SERIES MANUAL
13.1.10
V
T
This test measures the on-state voltage after the device has fired. Again, the device
must be fired using the programmed IBO current drive. After firing, the device is held
on by the 1 amp Low Source current drive for 6 to 8 mSecs while internal range
relays are set for the programmed I
T
current value. The I
T
pulse is then applied for a
period of 300
μSecs, at the end of which the V
T
measurement is made.
Figure 13-13 SSOVP V
T
TEST Circuit Diagram
Figure 13-14 SSOVP V
T
Test Waveform
13.1.11
V
Z
This is a zener test that provides measurement of the zener voltage of the device at
the programmed current level.
13.1.12
I
D
This is a leakage test that provides measurement of the device leakage current at the
programmed voltage level.
LO SRC
DRGT
1.2A
K15
RB
COM
RING
50
TIP
BO
I
I T
V FIRE
I
I
V
SET UP
BO
T
T
BO
Summary of Contents for 5000 Series
Page 2: ......
Page 4: ......
Page 10: ......
Page 30: ...TABLE OF CONTENTS xx Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 38: ...1 OVERVIEW 1 8 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 52: ...3 THEORY 3 6 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 62: ...4 SPECIFICATIONS 4 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 64: ...5 BASIC OPERATION 5 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 102: ...6 PC SOFTWARE 6 38 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 218: ...7 MAINTENANCE TROUBLESHOOTING 7 116 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 226: ...8 LOW CURRENT DECK 8 8 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 234: ...10 PIN PROGRAMMABLE SCANNER 10 4 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 242: ...APPENDICES 13 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 244: ...APPENDIX A 13 4 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 250: ...APPENDIX C 13 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 252: ...APPENDIX D 13 12 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 254: ...APPENDIX E 13 14 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 258: ...APPENDIX F 13 18 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 262: ...APPENDIX G 13 22 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 268: ...APPENDIX H 13 28 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 290: ...APPENDIX I 13 50 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 292: ...APPENDIX J 13 52 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 296: ...APPENDIX K 13 56 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 304: ...APPENDIX M 13 64 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 308: ...APPENDIX N 13 68 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 312: ...APPENDIX P 13 72 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 316: ...14 APPLICATION NOTES 14 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 318: ...APP 001 14 4 Version 1 23 MODEL 5000 SERIES MANUAL Figure 14 2 APP 001 Test Screen ...
Page 320: ...APP 001 14 6 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 322: ...APP 002 14 8 Version 1 23 MODEL 5000 SERIES MANUAL Figure 14 6 APP 002 Bin Sort Screen ...
Page 324: ...APP 003 14 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 326: ...APP 004 14 12 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 332: ...APP 006 14 18 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 336: ...APP 007 14 22 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 338: ...APP 008 14 24 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 340: ...APP 009 14 26 Version 1 23 MODEL 5000 SERIES MANUAL ...