8
– LOW CURRENT DECK
MODEL 5000 SERIES MANUAL
Version 1.23
8-5
8.4.2 5000 Series Tester Self-Test
Cable Connected Case:
To test the 5000 Series Tester, remove the Low Current Deck device cable from the
front panel of the 5000 Series Tester (it is not necessary to remove the Low Current
Deck control cable) and attach the 5000 Series Tester Self-Test Fixture as described
in Section 2.1.4. Set the “Remote Deck” switch (on the Low Current Deck panel) to
OFF and run the 5000 Series Tester self-test as described in Section 2.1.4.
Plug-In Case:
Run the Mainframe Self-Test with the Mainframe Self-Test fixture installed into the
Low Current Deck. If an error occurs, remove the Low Current Deck from the
mainframe, install the Mainframe Self-Test fixture into the mainframe and re-run the
mainframe self-test. If the error remains, then troubleshoot the mainframe as
described in Section 7. If the error clears then troubleshoot the Low Current Deck as
described below.
8.4.3 Lo Current Deck Self-Test
Cable Connected Case:
Reconnect the Low Current Deck device cable to the front panel of the 5000 Series
Tester. Set the “Remote Deck” switch to ON. Attach the Low Current Deck Self-Test
Fixture to the Low Current Deck panel jacks.
From the PCW software select the “Runtime” – “Self-Test” – “LCD Test” menu item.
Click the “Start” button. Successful Low Current Deck self-test will display “LODECK
OK”.
It is common after first receiving the Low Current Deck that an error code will result
from self-test due to the mercury relays not being settled. If an error occurs, confirm
all cable connections and switch positions and run the LCD Self-Test again.
Use the “Runtime” – “Self-Test” – “LCD Block” menu selection to run a specific block
for troubleshooting (see Section 8.4.3 for the Block error codes).
It is good practice to run both the 5000 Series Tester self-test and Low Current Deck
self-test at least on every system power up.
Plug-In Case:
Attach the Low Current Deck Self-Test Fixture to the Low Current Deck panel jacks.
From the PCW software select the “Runtime” – “Self-Test” – “LCD Test” menu item.
Click the “Start” button. Successful Low Current Deck self-test will display “LODECK
OK”.
It is common after first receiving the Low Current Deck that an error code will result
from self-test due to the mercury relays not being settled. If an error occurs, confirm
all cable connections and switch positions and run the LCD Self-Test again.
Use the “Runtime” – “Self-Test” – “LCD Block” menu selection to run a specific block
for troubleshooting (see Section 8.4.3 for the Block error codes).
Summary of Contents for 5000 Series
Page 2: ......
Page 4: ......
Page 10: ......
Page 30: ...TABLE OF CONTENTS xx Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 38: ...1 OVERVIEW 1 8 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 52: ...3 THEORY 3 6 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 62: ...4 SPECIFICATIONS 4 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 64: ...5 BASIC OPERATION 5 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 102: ...6 PC SOFTWARE 6 38 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 218: ...7 MAINTENANCE TROUBLESHOOTING 7 116 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 226: ...8 LOW CURRENT DECK 8 8 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 234: ...10 PIN PROGRAMMABLE SCANNER 10 4 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 242: ...APPENDICES 13 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 244: ...APPENDIX A 13 4 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 250: ...APPENDIX C 13 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 252: ...APPENDIX D 13 12 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 254: ...APPENDIX E 13 14 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 258: ...APPENDIX F 13 18 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 262: ...APPENDIX G 13 22 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 268: ...APPENDIX H 13 28 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 290: ...APPENDIX I 13 50 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 292: ...APPENDIX J 13 52 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 296: ...APPENDIX K 13 56 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 304: ...APPENDIX M 13 64 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 308: ...APPENDIX N 13 68 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 312: ...APPENDIX P 13 72 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 316: ...14 APPLICATION NOTES 14 2 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 318: ...APP 001 14 4 Version 1 23 MODEL 5000 SERIES MANUAL Figure 14 2 APP 001 Test Screen ...
Page 320: ...APP 001 14 6 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 322: ...APP 002 14 8 Version 1 23 MODEL 5000 SERIES MANUAL Figure 14 6 APP 002 Bin Sort Screen ...
Page 324: ...APP 003 14 10 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 326: ...APP 004 14 12 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 332: ...APP 006 14 18 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 336: ...APP 007 14 22 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 338: ...APP 008 14 24 Version 1 23 MODEL 5000 SERIES MANUAL ...
Page 340: ...APP 009 14 26 Version 1 23 MODEL 5000 SERIES MANUAL ...