Unwrapped Cycle
Sterilize light loads of solid metal instruments
such as hand instruments, speculums and
forceps.
2000 G4:
This is NOT a sterilization cycle. Refer to
Operator’s Manual.
5000 G4:
Sterilize larger loads of unwrapped metal
instruments.
Drying PerformanCe
four steps to obtaining optimal drying results
After cleaning and drying the
cassette, treat the interior
surfaces with STAT-DRI
™
Plus.
Coat Cassette
Use Drying Plate rack
(5000 only)
For optimal drying performance of wrapped
instruments, use the drying plate rack
supplied and insert one bag per plate
with window side up.
Set Up Waste Bottle
Fill this bottle with water to
the minimum level and set it
on, or close to the floor.
Plate positions
for full load
(10 bags).
With thumbs on the thumb
pads, push down, pulling up
with forefingers until lid opens.
opening / Closing
the Cassette
®
Hold the cassette handle in one
hand and the carry handle in the
other. Place the end of the cassette
into the unit and drop the carry
handle into its closed position.
inserting Cassette
filling Cassette
5000 model
2000 model
5000 model
Gently push cassette inward until
you hear a click. Never force the
cassette into the STATim.
2000 and 5000 models
Cycle Starting guide
1
2
3
4
1.
2. Scroll to Level and select.
3. Adjust the leveler
feet to move the
bubble. Level
will change from
red to green.
Levelling your Unit
•
Replace water as needed.
•
For ophthalmic use, drain at the
end of every workday, leave empty,
and refill at the start of the next
workday.
•
Empty the waste bottle every time
you refill the water reservoir.
•
Fill the bottle with water, up to the
MIN line marking. You may also add
some chlorine-free disinfectant.
•
Wash the interior of the cassette with dishwashing soap or a
mild detergent that does not contain chlorine.
•
Scrub the inside with a cleaning pad designed for use with
Teflon
®
-coated surfaces.
•
After removing all traces of the detergent, treat interior sur-
faces of the cassette with the STAT-DRI
™
Plus drying agent to
enhance drying process. Additional STAT-DRI
™
Plus can be
ordered by requesting 2OZPLUS, 8OZPLUST, or 32OZPL US.
•
Check the water reservoir filter every week and clean
if necessary. Replace only if necessary.
•
Replace every 500 cycles or
six months (whichever is first).
Call for service if wet.
•
Replace every 500 cycles or
six months (whichever is first).
•
Replace every 500 cycles or
six months (whichever is first),
or whenever necessary.
Cassette
Water Reservoir
Filter
Biological Filter
Air Filter
Cassette Seal
Water
Reservoir
Waste
Bottle
toUCh SCreen menU oPtionS
network Connectivity –
Indicates you are connected to
the network.
USB Storage Device –
Access last 5 cycle data and
last 5 cycle faults. Indicator also
confirms proper USB connection.
Water Level / Quality Status** –
Indicator will notify user if water
is of sufficient quality or quantity
to proceed with sterilization.
Cassette Status** – Indicator
will notify user if cassette has
been properly inserted.
User/technician menu – Access
unit information and menus to
change unit settings.
Start Button – Button is
enabled once a cycle type has
been selected and used to begin
a cycle.
StoP Button – Button is enabled
after a cycle has started and used
to interrupt a cycle.
Toggle
UP/DoWn to
view menu options.
enter
Push button to view/select
menu options.
BaCK
Push button to return
to previous menu.
Wrapped Cycle
Sterilize solid or hollow metal instruments which
have been sealed in autoclave bags.
rubber and Plastics Cycle
air Dry only Cycle (2000 g4)
heavy Duty Unwrapped Cycle (5000 g4)
Sterilize solid or hollow unwrapped instruments.
DaiLy
Do not overload cassette.
StatIM
generaL USage
Operator’s Manual should be consulted before using this autoclave
** NOTE: Sterilization cycle will not begin if there is an error
USer/teChniCian
menU anD navigation
generaL maintenanCe
to ensure trouble-free performance, follow this preventative maintenance schedule.
WeeKLy
every 6 monthS
www.
scican
.com
STATim is a registered trademark and STAT-DRI, Dri-Tec
R
and Dri-Tec
C
are trademarks of SciCan Ltd. Teflon is a registered trademark of E.I. Du Pont de Nemours and Company Corporation.
95-112742 Rev. 2.0
Best Practices: SciCan recommends the final user carefully
choose the most appropriate sterilization cycle according
to the recommendations of their leading infection control
authorities and local regulatory guidelines / recommendations.
09-110755 Rev1: Artwork, Dri-Tec-C
Designed: March 11, 2008
PANTONE 363 C
BLACK
TEXT FONTS:
Dri-Tec: Frutiger Bold
STEAM HEAT DRYING: Copperplate Bold
20% Black
Outline around Dri-Tec
09-110757 Rev1: Artwork, Dri-Tec-R
Designed: March 11, 2008
PANTONE 280 C
20% Black
Outline around Dri-Tec
BLACK
TEXT FONTS:
Dri-Tec: Frutiger Bold
STEAM HEAT DRYING: Copperplate Bold
2000 and
5000 models
2000 model
5000
model
4
8
4
8
4
1
2
3
4
1
2
3
4
1
2
3
4
4