Please read the Operator Manual and this Quick Reference Guide before using your Aquastat Water Distiller.
1. Start-Up
Place your Aquastat on a dry, level surface in a well-ventilated area. Plug the cord into a properly grounded
outlet.
2. Cleaning Your Aquastat
1. To clean the water bottle, tighten the reservoirvalve and fill the
water bottle 3/4 full with water.
2. Add 1-2 drops of dishwashing detergent and vigorously shake the
water bottle.
3. Empty the bottle and thoroughly rinse to remove detergent.
3. Sanitizing Your Aquastat
1. Add 5 ounces (150Ml) of 3% hydrogen peroxide solution (H
2
O
2
) to the water bottle.
2. Fill the bottle with hot (160
°
F/71
°C
) distilled or tap water.
3. Cap the water bottle and shake vigorously.
4. Let stand for twenty minutes then drain thoroughly.
5.Rinse with room temperature distilled water, then drain. The
water bottle may be used immediately or allowed to air dry.
4. Fill the boiling chamber.
1.Remove the boiling chamber from the chassis and remove
the boiling chamber cap. Note the four tabs at the base of the boiling chamber's fill opening.
2.Fill the boiling chamber with cold tap water up to the bottom of the four tabs.
AQUASTAT
Quick Reference Guide
Boiling Chamber Cap
Tabs
Baffle
Start Button
Cord
Cooling Vents
Baffle
Baffle Gasket
Boiling Chamber
Water Bottle
Reservoir
Valve
Cap
Boiling Chamber Cap
Distiller Chassis
Baffle Nut
O-Ring
Dispensing
Valve
Boiling
Chamber
Chassis
Gasket