ProLab 4000
Ion concentration
97
ba75646e03
12/2012
9
Ion concentration
9.1
General information
Temperature
measurement with ion
selective measurements
The temperature value is documented with all measurements.
You have the following options to measure the temperature:
Automatic temperature measurement with
– the temperature sensor integrated in the electrode (NTC30 or
Pt1000).
– the integrated temperature sensor of another electrode that is
connected and immersed in the sample at the same time.
– an external temperature sensor (NTC30 or Pt1000).
Manual determination and input of the temperature.
The display of the temperature indicates the active temperature mea-
suring mode:
Operation with burettes
Measurements with ISE methods can be automated by controlling
burettes with the meter:
Connect the burette to the RS232 interface of the ProLab 4000
If necessary, interlink the burettes
(see also operating manual of the burette).
Activate the burette control on the ProLab 4000:
Measuring
/
Burettes settings...
/
Activate burettes
(
)
Set the address of the burette on the meter and on the burette
(see page 102 and operating manual of the burette).
Note
Incorrect calibration of ion sensitive electrodes will result in incorrect
measured values. Calibrate regularly before measuring.
Temperature
sensor
Resolution of the
temp. display
Mode
yes
0.1 °C
Automatic with tempera-
ture sensor
-
1 °C
Manual
Note
More information on automation with burette control is available on the
Internet under www.si-analytics.com.
Summary of Contents for ProLab 4000
Page 4: ...ProLab 4000 4 ba75646e03 12 2012...
Page 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Page 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Page 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Page 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Page 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Page 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Page 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Page 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Page 227: ......