ProLab 4000
System functions
63
ba75646e03
12/2012
6.
8
Automatic stability control
The
Autom. stability control
function continuously checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values.
You can activate or switch off the
Autom. stability control
function
(menu,
Autom. stability controlSystem
).
The
Autom. stability control
function is carried out:
as soon as the measured value is outside the allowed stability range
when you switch over between the measured parameters with
<MODE>
.
The stability of the measured value is indicated by the background
color of the measured value. In the delivery condition, the SC stable /
SC not stable conditions are assigned to the following colors:
Color
Meaning
red
SC not stable
Stability criteria not met
green
SC stable
Stability criteria met
SC stable
SC not stable