Cold light sources are devices for providing intensive illumination.
The infrared (thermal) components of the bulb output are filtered out.
The intensive visible light is transmitted to the object to be illuminated through flexible or
moveable self-supporting light guides.
1. Important notes
Please take particular care that
your KL 200 cold light source is supplied with the voltage stated on the nameplate
all ventilation slits are kept free (base and upper surface of device);
the bulb has cooled down before it is replaced. As a reminder, a warning sign
is
attached to the cover of the bulb compartment.
The cold light source has been designed for operation in dry rooms.
The only way to ensure that the power supply has been disconnected completely is to
unplug the power plug from the mains.
The device is not suitable for use in areas where there is a danger of explosion.
Repairs may only be carried out by the manufacturer or by persons authorised by the
manufacturer.
Please read these instructions carefully and follow them. No guarantee can be given on the
safety of the device in the case of failure to comply with the instructions given.
The manufacturer accepts no liability for damage resulting from failure to comply with the
instructions given here.
!
On/off switch with three position light
level switch
Light guide socket
Ventilation slots
Nameplate on base of lamp
Mains cable
(with UK plug for UK market)
Bulb compartment with halogen
cold-light reflector bulb
Fuse accessed via base of lamp
1
2
3
4
5
6
7
4
2. Operation
2.1. Installing a light guide
Push the light guide into its mounting socket
as far as it will go; in the case of light guides with
a guide pin take care that this pin locks into one
of the slots.
Installing a light guide
On/off switch
3. Replacing the bulb
Open the bulb compartment as far as it will go pulling gently on the handle recess. When
the assembly swings out, the power supply to the lamp is automatically interrupted. Allow the
defective bulb to cool down; then remove it; insert the new bulb; close the bulb compartment;
the flap will lock into place easily.
Replacing the bulb
2.2. Starting up operation
On/off switch :
Position 0
Lamp is switched off
Position I
Bulb is operated at reduced voltage
and Position II
This extends its service life considerably
Position III
Bulb is operated at full voltage
Colour temperature approx. 3200 K
Average service life of bulb is 300 h
1
2
6