ProLab 2000
What to do if...
99
ba75577e03
11/2009
CalClock flashes
6.2
ISE measurement
Error message
Err1
Obviously incorrect
measured values
Error message,
Error
(invalid
calibration)
– Measurement procedure not
suitable
– Follow special procedure
Cause
Remedy
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Measuring range exceeded
– Dilute test sample
Cause
Remedy
– Electrode not connected
– Connect electrode
– Cable broken
– Replace electrode
Cause
Remedy
ISE electrode:
– Moisture in the plug
– Dry plug
– Electrode obsolete
– Replace electrode
– Electrode unsuitable for the
range to be measured
– Use suitable electrode
– Socket damp
– Dry socket
Calibration procedure:
– Wrong sequence of standards
for three point calibration
– Select correct sequence
– Calibration standards do not
have the correct temperature
(max.
±
2 °C temperature
difference)
– Adjust the temperature of
the calibration standards