MAGLIFE light
Part number: 0-48-0080
Page 5
WARNINGS, PRECAUTIONS AND NOTES
Please read and adhere to the following list of warnings, precautions and notes. Some of them have been
repeated at appropriate areas throughout this manual.
Follow the warnings and precautions stated on the labels affixed on the monitor. Throughout this manual we
have added specific notifications that offer additional information. These notations are designated as:
A
Note
is provided when extra general information is applicable.
A
Caution
is provided when special care is to be exercised by the user and/or patient, to avoid injury to the
patient, damage to the device or damage to other property.
A
Warning
is provided when actions may result in a serious outcome (i.e. injury, serious adverse effect,
death) to the patient or user.
Caution: MAGLIFE
light
is a monitor designed solely for use close to Magnetic
Resonance Imagers of 0.2 to 3 T for monitoring of patients undergoing an MRI
examination.
Caution:
The continuous presence of a qualified person is imperative throughout the
examination.
Warning:
Before using the
MAGLIFE light
monitor, follow the safety instructions below:
MAGLIFE light
may only be used by trained healthcare workers who are
familiar with its instructions for use.
Make sure that the voltage and frequency of the electricity system match
those stated on the identification plate.
Warning:
Because the device is a class I device, it may only be used in premises with
an electrical system comprising an earth connection.
Warning:
Connecting other devices or sensors to the patient could lead to leakage
currents that may be harmful to the patient. Consult SCHILLER before
interconnecting with other equipment.
Warning:
Take account of the physiological effect of the other devices connected to the
patient.
Warning:
This device may not be installed or operated in explosive environment.
Warning:
This device is not designed for use with inflammable anaesthetic agents.
Caution:
Use only sensors supplied by SCHILLER. They have been designed specially
for use in MRI environments.
Warning:
MAGLIFE light
is designed for use with MRI systems with a magnetic field
between 0.2 and 3 Tesla. Contact SCHILLER for use beyond that magnetic
field range.
Warning:
It is imperative the
MAGLIFE light
be installed in the area around the magnet
where the magnetic leakage field is less than or equal to 40 mT (400 Gauss)
(see typical installation drawings and labels on the device).
Summary of Contents for MAGLIFE LIGHT
Page 1: ...MAGLIFE light User manual Version 1 2 October 2006 Part No 0 48 0080 ...
Page 39: ...MAGLIFE light 4 8 Structure of menus 4 8 1 User menus Part number 0 48 0080 Page 39 ...
Page 40: ...MAGLIFE light Part number 0 48 0080 Page 40 ...
Page 50: ...11 Front 5 1 3 4 2 6 7 8 9 10 11 12 Part number 0 48 0080 Page 50 ...
Page 51: ...12 Rear and power supply 13 15 16 17 18 14 19 20 Part number 0 48 0080 Page 51 ...