Page 123
Maintenance
10
User guide
Inspection and Checklist Tables
10.8
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
10.8.2
Every 12 months
10.8.3
Lifed-item replacement every 5 - 10 years
Inspection
Results
Inspection
Functional, safety checks and in-
spections
Confirm the date of the last
factory inspections and tests
• Return the unit to your nearest
authorised service point or your
SCHILLER agent for safety and
functional checks.
o
o
o
o
o
Date of inspection:
Inspector:
Inspection
Results
Replacement
Battery
Replace battery
• The battery needs to be replaced:
– when the operating time is less
than 1 hour, see section
– Replace the safety primary cell
(see expiring date or when de-
vice switches of immediately
when replacing power battery)
– Replace the internal button pri-
mary cell (every 10 years)
o
o
o
o
o
o
o
o
o
o
o
o
o
o
o
Date of replacement:
Inspector:
Defibrillation capacitor
Replace defibrillation capacitor
• Send the unit to your nearest
SCHILLER service centre for
capacitor replacement if the
defibrillation capacitor deviates
more than 15 % [joule] from the
intended value.
o
o
o
o
o
Date of replacement:
Inspector: