Page 43
Care & Maintenance
5
A"#$0$
Service Handbook
Maintenance interval for the battery
5.4
A
rt
.-
no
.:
2
.5
4
0
0
37
R
ev
.:
a
5.4 Maintenance interval for the battery
• during normal operation no maintenance necessary.
• If not used every 3 months.
• replace the battery approx. every 4 years (depending upon application) if the actual
running time falls substantially under 1 hour.
5.4.1
Charging the battery
5.4.2
Testing battery
1. Connect ECG simulator.
2. Switch on the device (without connecting to the mains supply)
3. Select Menu SETUP/SYSTEM SETTING/ TEST AND INFO and enter following
code: 032.
A full test print out will be executed. The device must run 4 hours when the battery
was fully charged.
Important
The battery is maintenance free during its normal life.
The battery should remain charged during storage. If the storage period exceeds
three months, recharge the battery.
Philips screwdriver 1
Important
A totally discharged battery requires approx. 7 hours to be 90% recharged.
It is possible to use the unit when the battery is being charged. However, when this is
the case, the charging time of the battery will be substantially extended!
1. Connect the device to the mains but do not switch it on.
2. The LED for mains supply (1) is lit.
3. Charge the battery for at least 7 hours.
1
Summary of Contents for AT-101
Page 1: ...A 0 ervi e andboo3 Art no 7 5400 Rev a 400 A 0 7 annel E G Devi e ...
Page 6: ...Page 4 Art no 2 540037 Rev a ...
Page 42: ...5 Care Maintenance 5 2 Functional test A 0 Page 40 Art no 2 540037 Rev a ...
Page 63: ...Page 61 Appendix 8 A 0 Service Handbook Checklist 8 1 Art no 2 540037 Rev a Other remarks ...
Page 64: ...8 Appendix 8 1 Checklist A 0 Page 62 Art no 2 540037 Rev a ...