Disassembly Guide
ARGUS VCM
Monitor Disassembly
This section describes the steps to separate the front and rear case assemblies.
Figure 17. Monitor Disassembly
Table 11. Part Descriptions – Monitor Assembly
Part Codes
Descriptions
Qty
-
Front case assembly (A)
1
-
Rear case assembly (B)
1
-
Temperature case assembly (C)
1
S8071-0
Screw 3 × 8 (tapping)
2
S8070-0
Screw 3 × 8 (flat-head)
3
S8091-0
Screw 3 × 12
3
S8092-0
Screw 3 × 16
2
Before the steps A, B and C
1. If Temperature option installed, remove 2 tapping screws (S8071-0) on the rear of the
temperature case.
2. Disconnect Temperature wire to separate the temperature case from the rear case.
3. Remove 3 flat-head screws (S8070-0) on the bottom of the monitor.
4. Remove 5 screws (S8091-0, S8092-0) on the rear case.
5. Separate the front case from the rear case.
6. Carefully disconnect all cables running between cases: disconnect SMPS wire, NIBP
cable, Printer wire, Speaker wire, Serial wire and NIBP hose.