Summary of Contents for Proscan 2200

Page 1: ...00 User Manual Scantron Industrial Products Ltd Monarch Centre Venture Way Telephone 44 0 1823 333343 Taunton Fax 44 0 1823 333684 TA2 8DE e mail scantron scantronltd co uk England Internet www scantronltd co uk ...

Page 2: ...r Manual V 4 0 Documentation Version Version Created Amended Author Notes 4 0 03 06 2020 JFH Components Installed Supplied Notes Sensor Type Stil Chromatic Sensor Controller Stil Prima CCS Sensor Display 7 Colour Camera USB Firewire CMOS ...

Page 3: ...e rear of the PSU Module 11 3 4 Connections to the rear of the stand 11 3 5 Main Power Connection 12 3 6 Switching On 12 3 7 Switching Off 13 4 0 Changing the sensor 14 4 1 Removing the Nose Cone 14 4 1 1 Changing S Type Sensors 15 4 1 2 Sensor Mounting 15 5 0 Dark Referencing 16 5 1 Acquiring the dark signal 16 5 2 Recommendation 16 6 0 Manual Z Height Control Buttons 17 7 0 Camera 17 7 1 Camera ...

Page 4: ...rom medical and dental materials and ceramics to rough paper The Proscan can be equipped to cater for virtually any measuring need 1 2 Options To accommodate clients specific needs this manual may include references to sensors cameras and display interfaces that have not been supplied installed on a particular system The sensors supplied with the Proscan 2200 system may incorporate one of two main...

Page 5: ...technology low mechanical noise Extremely high positional accuracy and repeatability Inaudible in operation Real time electronics Accurate Reliable Flexible The Proscan 2200 Profilometer consists of a stand containing the precision x y z table motion control electronics and a proprietary communication control and data acquisition control board designed by Scantron Artemis A compact power supply mo...

Page 6: ...c Sensor Controller Technology Chromatic Confocal Working frequency Up to 2 KHz Sensitivity High Compatible optical pen CL MG OP Everest and PROBE Overall dimensions 430 mm x 350 mm x 140 mm max Weight 6 1 kg Electrical Specification 100 110 or 220 240 VAC 50 60 Hz see product labelling ...

Page 7: ... optical pen or optical probe is the measuring head of the sensor It is connected to the controller via an optical fibre The sheath is generally orange in colour The various optical pens are defined by different measuring ranges and resolutions Each optical pen offers a working distance a measurement dynamics a spot size on the target as well as an angular opening of the light beam making it uniqu...

Page 8: ...such it is important that consideration is given to its location and setup In addition the stand also houses custom communication control and data collection ARM PCB Artemis board and power electronic components Subsequently it should not be placed near areas of extremely strong magnetic fields The stand should be mounted on a very robust vibration free rigid surface Do not place on an anti vibrat...

Page 9: ...stem is scanning a line and is acquiring data POWER ON indicator Module is powered and switched on green on 240 VAC red on 120 VAC systems 5 V LED Indicates that the 5 VDC power supply is providing power 24 V LED Indicates that the 24 VDC power supply is providing power Back panel POWER ON switch Top depressed PSU Module is switched off Bottom depressed PSU Module is switched on 2 4 Power Requirem...

Page 10: ... The system is supplied with all communication cables These include RS232 Sensor communication between Proscan and CCS controller USB or Firewire between Proscan and PC USB between CCS Controller and PC Fibre Optic cable between sensor pen and CCS controller Care should be taken when placing or routing the fibre optic cable not to kink or bend the cable in tight curves ...

Page 11: ... damage any of the pins by forcing the connector Connect an IEC power cable to the socket and after checking that everything is safe and that all cables are correctly connected at both ends the PSU may be switched on using the rocker switch next to the IEC connector 3 4 Connections to the rear of the stand Only connect disconnect cables with the system PSU Module fully powered down First connect t...

Page 12: ...l Number label 3 6 Switching On Keep fingers and objects clear of the stand when powering the system for the first time Once all the cables are in place and all objects are clear of the stand it is safe to switch the system on On switching on the PSU module the 5 and 24V LED indicators on the front of the Module should illuminate Correct booting of the Artemis board is indicated by the regular fla...

Page 13: ...ts home position when it will stop Simultaneously the Y axis of the stage will make a small move in a positive direction away from home in the bottom right hand corner It will then move in a negative direction towards home until it finds its home position when it will stop The full software screen will be displayed Please see the Software Manual for full information on operation of the software 3 ...

Page 14: ...cone on each side and lift towards you and upwards off the machine The nose cone rests on two rods which protrude from the z drive these locate in the slots on the rear of the nose cone Slots Locating rods Nose cone Z Drive Replacement of the nose cone is the reverse of removal Please note the camera illustrated my differ from that installed Keep fingers and objects clear of moving parts ...

Page 15: ...scan ensure that the correct sensor is selected on the S type control unit Select the correct sensor in the Proscan 2200 software Before taking measurements allow the sensor to Warm up fully approx 15 minutes Dark reference the sensor Ensure unused sensors are stored where they not going to get damaged 4 1 2 Sensor Mounting The sensors are fitted onto a mounting plate which locates onto the front ...

Page 16: ...urement may be launched by pressing the Dark button on the sensor front panel This operation may take a few dozens of seconds as the sensor measures and saves the dark signal at all pre set frequencies successively In order to perform a dark signal acquisition it is essential to have no object within the measurement field or even better to blank off the light beam by applying a piece of paper over...

Page 17: ...on simultaneously with the Up Down button will set a higher move speed for initial placement The final position can be fine tuned at the lower speed The sensor display is configured to show the percentage of the range rather than height reading as the mid range point would differ for a particular sensor pen 7 0 Camera The Proscan 2200 system may have optionally a camera installed to aid correct pl...

Page 18: ...optics are clean and free from dust Clean using a lint free wipe and a small amount of methanol Every 3 months after handling if data quality falls X and Y axis linear encoder strip and readhead Use a clean dry lint free cloth Every 12 months if readhead shows red Use only the following solvents sparingly with a wetted lint free cloth Propan 2 ol iso propyl alcohol CH3CHOHCH3 N Heptane CH3 CH2 5CH...

Page 19: ...le If any of the indicators are not lit contact Scantron Software displays message The scan only contains invalid data and cannot be displayed Wrong sensor selected i e sensor on system is not selected in software Go to Tools Sensor Selection and chose the correct sensor Also ensure sensor is selected on the S type controller if applicable Sensor controller not switched on S type Turn sensor contr...

Page 20: ...can is required Error message setup does not lie within stage area 100 00mm 100 00mm Please check your settings in new scan mode Stage has a fixed size so cannot scan areas larger than 100mm by 100mm Check that your scan parameters are not going to take you outside of the stage range from your start position No picture in camera software Camera not in focus Rotate focusing ring on camera to adjust...

Page 21: ...ment control and laboratory use EMC requirements General requirements BS EN 61326 2 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Particular requirements Test configurations operational conditions and performance criteria for sensitive test and measurement equipment for EMC unprotected applications BS EN ISO 12100 2010 Safety of machinery General principle...

Page 22: ...4 0 10 2 Contacting Scantron Postal Address SCANTRON INDUSTRIAL PRODUCTS LTD Monarch Centre Venture Way TAUNTON Somerset TA2 8DE England Telephone 44 0 1823 333343 Fax 44 0 1823 333684 e mail scantron scantronltd co uk Internet www scantronltd co uk ...

Page 23: ...will despatch spare parts and service personnel to the buyers facility to carry out the repair at its own expense The buyer will cover the shipping costs of the parts and the travelling and accommodation costs of the personnel The buyer will leave the defective equipment at our disposal for a period no longer than 1 month to complete the repair The warranty does not cover damages caused by careles...

Page 24: ...ive and complete with quick change fibre optic cable Labelled according to sensor type Varies 13 IEC Power lead 1 14 3 m serial cable with 15 way M and 9 way M D connectors 15 way end labelled SENSOR connects to the stand 9 way end labelled SENSOR connects to the sensor controller 1 11 3 L type Laser Sensors 15 L Type Laser Sensor Controller Contained within the PSU Module Not visible from outside...