RTC
®
5 PC Interface Board
Rev. 1.9 e
10 Commands And Functions
485
innovators for industry
Normal
List Command
set_softstart_level_list
Function
similar to
(see above), but a list command
Call
set_softstart_level_list( Index, Level1, Level2, Level3 )
Parameters
See
Comments
• The
command defines three
Level
values for the pulses number
Index
through
Index+3
. The command must be called separately for each individual
Level1/2/3
value-triple.
•
Index
must be in the range [0…1023], otherwise the
set_softstart_level_list
command will, already during loading, be replaced by a
return
code
RTC5_PARAM_ERROR
).
• Softstart won’t use level values where the index count exceeds
Number
(
Number
is spec-
ified via
).
• Otherwise this command is identical to the control command
(also
see the comments there).
• If the “freely defined wobbel shape“ wobbel mode is activated, then this command will
chapter 8.4 "Wobbel Mode", page 167
.
RTC
®
4
RTC
®
5 new command
In RTC
®
4 compatibility mode: see
Version info
Available as of version DLL 536, OUT 536
References
,