7.4 Laser Control
The RTC
®
5 provides several laser control signals
with programmable pulse length and frequency via
the LASERON, LASER1 and LASER2 ports of the 15-pin
D-SUB-laser connector (see
). These signals
can be used for controlling various types of lasers.
,
the user can choose one of seven different laser
control modes:
• The CO
2
mode (laser mode 0) is provided for
controlling CO
2
lasers.
• The four YAG modes (laser modes 1, 2, 3 and 5)
are provided for controlling Nd:YAG and related
lasers.
• Laser mode 4 and laser mode 6 generate a contin-
uously-running control signal.
Depending on the selected mode, different laser
control signals are available for “laser active” oper-
ation and “laser standby” operation. The various laser
modes are described in the following sections.
All laser control signals are TTL signals. They can be
set to either active-high or active-low via the
command
– also
see
. The command
(Bit#13) can be used to query whether the signals are
currently set to active-high or active-low.
The maximum current load for all signals is 20 mA.
7.4.1 Enabling, Activating and
Switching Laser Control
Signals
After a hardware reset and also after initialization
with
, all laser control signals are
disabled. All laser output ports (LASERON, LASER1
and LASER2) are in the (high impedance) tristate
mode. TTL states (LOW or HIGH) will only be available
if the signals are enabled via the
command (and the desired TTL level is thereby spec-
ified).
After
is called:
• Laser control signals for “laser active” operation
will be provided at all output ports as soon as
these signals have been
activated and switched
on
or
• All output ports will provide standby signals if
they have been activated and if no laser control
signals for “laser active” operation are switched
on
or
• The outputs will be in their basic TTL states (e.g.
LOW if a high signal level was specified) if no laser
control signals (neither for “laser active” nor for
“laser standby” operation) have been activated
After
output ports (LASERON, LASER1 and LASER2) will
enter the (high impedance) tristate mode only after a
new hardware reset or if the laser control signals are
subsequently disabled via a command (see
,
(bit#2 = 1)).
Before activation of the laser control signals, the
command
should be called for
selecting the laser control mode (otherwise – by
default – the CO
2
mode is selected).
For first-time activation of the laser control signals
(both for “laser active” and for “laser standby” oper-
ation), the
command must be
used to enable the laser output ports (LASERON,
LASER1 and LASER2).