System Modes
System operation can take on one of several modes as defined by interior mode selection controls
located on the Main Processor Board. A two-position, Mode Slider Switch and a three-pin, Setup Mode
Jumper field are used to configure the mode as previously described in section titled
. Any run-time changes made to the mode selection controls are evaluated and acted upon
whenever the user navigates through the main menu options by way of the Keypad or Rotary Knob.
When such a mode change is detected, a temporary message will be displayed indicating the new mode.
Each mode configuration offers the user a unique set of control menu options as detailed in section
titled
7.1
Run Mode
This is the normal operational mode as a measurement system and offers a minimal set of control
options to the end-user. It is selected by moving the Mode Slider Switch into the Run Mode position,
which is when the
text “RUN”
is visible beneath the slider. The configuration of the Setup Mode Jumper
has no effect while in Run Mode.
7.2
Setup Mode
This mode enables additional control options depending on a specific purpose. It is selected by moving
the Mode Slider Switch into the Setup Mode position, which is when the text
“SETUP”
is visible beneath
the slider. A more specific setup mode is further selected via placement of a shorting-shunt on a 3-pin,
Setup Mode Jumper field.
•
User Setup Mode
: This is a high-level setup mode that offers the end-user limited access to system
menu options that augment the normal operation of the system. It is selected by installing a
shorting-shunt on the 3-pin, Setup Mode Jumper such that the text
“USER” remains visible
within
the jumper field.
•
Factory Setup Mode
: This is a low-level setup mode that offers a factory technician greater access
to system menu options needed during system assembly and calibration. It is selected by installing
a shorting-shunt on the 3-pin, Setup Mode Jumper such that the text
“FACTORY”
remains visible
within the jumper field.
If the Mode Slider Switch is in the SETUP position, a
shorting-shunt MUST be placed on the Setup Mode
Jumper to further select the setup mode. The absence of
a shorting-shunt will keep system operation in Run Mode
.
NOTE: