58
SBS-3510 User’s Manual
15 Troubleshooting
Table 7: Error messages
Error message
Cause
Out of Specification The measured sample temperature is outside the
specifications.
Out of Range
The measured value is outside the specified range.
Possible reasons:
• The temperature is too high or too low.
• The density is too high or too low.
• The measured value is invalid.
Temperature Range The temperature of the reference liquid for
readjustment is outside the temperature
specifications
(+15°C to +25°C).
Density Range
The density of deionized water at 20°C measured
during readjustment is outside the allowed limits
(tolerance 0.01 g/cm
3
).
Criterion
The density of deionized water at 20°C measured
during readjustment deviates from factory
adjustment by more than 0.01 g/cm
3
.
No Oscillation
The measuring cell is not oscillating.
Possible reasons:
• The measuring cell is only partially filled:
Fill the sample again.
• The measuring cell is defective.
Sensor Error
The sensor in the measuring cell is defective.