SBS-3510 User’s Manual
9.3 Performing a Measurement
1. Perform all checks in section 9.1.
2. Fill the sample, see section 9.2.1 for filling with the filling tube or
section 9.2.2 for filling with a syringe).
The instrument shows continuous live readings of measuring values.
3. Start the measurement:
- Tap the
key,
- or use the corresponding gesture (see section 5.7) if gesture control
is enabled.
You may abort a measurement by tapping <Cancel> or with the
corresponding gesture if gesture control is enabled.
4. When the measurement is finished (indicated in the status bar), the
measuring result will “freeze” on the display, and the result is stored in
the internal data memory.
Tap <Live> to “unfreeze” the measuring result and return to showing
continuous live readings.
10 Measurement/Check Data in
the Data Memory
Measurement data as well as check data are stored automatically in the data
memory.
The data memory can hold a total of 1024 data sets. Further measurements
will overwrite the oldest stored data sets.
All stored data sets include the date and time of the measurement or check.
10.1 Viewing Data
1. In the quick access area, activate
(“Data” function).
The last data set will be displayed
2. Use the arrow keys to browse through the data sets in the data memory.
The index of the currently displayed data set is shown in the header
beside the total number of data sets stored.
Measurement data are shown with gray background, check data with
blue background.
3. Tap <Back> to return to the main screen.