SBS-2510 Digital Density Meter
Meter
17
2 Mov e the nav igation joy stick upwards and let go.
The sample pump starts and stops when the
piston is in the uppermost position.
3 Lif t the f illing tube out of the cleaning solution and
hold it ov er the waste beaker (2).
4 Mov e the nav igation joy stick downwards and let go.
The sample pump starts and stops when the
piston is in the lowermost position.
5 Clean the end of the f illing tube with a clean tissue.
6 Repeat the steps until all residue f rom the
sample is remov ed.
7 Mov e the nav igation joy stick upwards and let go.
The sample pump f ills the measuring cell with air.
8 Mov e the nav igation joy stick downwards and let go.
Most of the cleaning solution is remov ed
f rom the measuring cell.
9 Clean the end of the f illing tube with a clean tissue.
7.3 Check the m easurement accuracy
To check the measurement accuracy of the SBS-2510, y ou need to conf igure and
perf orm a test
7.3.1 Example: Test w ith w ater
The f ollowing chapters show y ou how to conf igure and perf orm a guided test with deionized
water at room temperature 77.0°F (25 °C).
7.3.1.1 Configure the test
The density unit is set to [g/cm
3
].
Confirmation of results
in
Settings
is activ ated.
1 Press the
Menu
key .
2 Nav igate to
Test
.
The
Test
window opens.
3 Change the parameter settings to the v alues shown in the f ollowing table.
Parameter
Setting
Explanation
Workflow
Guided
The SBS-2510 prompts users to f ill and drain
the measuring cell.
Standard type
Water
The SBS-2510 uses predef ined v alues f or the
water measurement.
Standard ID entry
Variable
The SBS-2510 prompts users to enter the
identif ication bef ore the density is measured.
Fill cycle
3
The SBS-2510 prompts users to f ill and drain the
measuring cell twice and then f ill the measuring
cell f or the measurement.
Fill speed
High
The pump f ills the measuring cell with 80 %
of the maximum pump speed.