SK-110TRH CALIBRATION

 

 
 

To calibrate the thermohygrometer, use a temperature/humidity  generating chamber to 

produce reference temperature and reference humidity. A chamber capable of achieving a 

temperature repeatability within ±0.1°C, humidity repeatability within ±1.0%, and humidity 

variability between 0 and 90% is required.

 

 
 

Equipment 

 

• 110TRH 

• 

Temperature/humidity generating chamber 

• Reference 

thermometer 

• Torque 

screwdriver 

• 

Phillips head screwdriver 

• 

Flat head screwdriver 

 

 

1. 

Remove the six screws from the device as shown. 

 

2. 

Remove the front section.

 

 
3. 

Insert four batteries into the case in the directions indicated.

 

 
4. 

Press and hold the POWER key for more than one second to activate the device.

 

 
5. 

Set the sensor in the temperature/humidity generating chamber and allow it to adjust to 
the temperature and humidity.

 

 
6. 

Temperature and humidity calibration 

 