Operating Instructions

Sartorius Extend Series
Sartorius Gem and Gold Extend Series

Electronic Analytical and Precision Balances and Precious Metal Scales

98648-013-98

Summary of Contents for ED124S-0CE

Page 1: ...Operating Instructions Sartorius Extend Series Sartorius Gem and Gold Extend Series Electronic Analytical and Precision Balances and Precious Metal Scales 98648 013 98...

Page 2: ...2...

Page 3: ...Weighing in Percent 29 Calculation 31 Animal Weighing Averaging 33 Net total Formulation 35 Totalizing 37 Mass Unit Conversion 39 Density Determination 41 ISO GLP compliant Printout Record 44 Data Int...

Page 4: ...th extreme caution when using pre wired RS 232 connecting cables as the pin assignments may not be compatible with Sartorius equipment Before connecting the cable check all pin assignments against the...

Page 5: ...supplied with models GK3102 GK2202 GW6202 0CEDS GW3202 0CEDS Metal disk Draft shield chamber with sliding doors Additional equipment supplied with models ED623S CW ED523S PCE ED423S CW ED323S CW ED153...

Page 6: ...n the components listed below in the order given Place the lower lid on the balance scale with the raised edge facing upwards and turn it until it is firmly in position Pan support Weighing pan Glass...

Page 7: ...e power cord into the wall outlet mains Use an original Sartorius AC adapter with a wide input voltage range 100 to 240 V order no 6971966 and replaceable power cord 6900900 Europe 6900901 US CDN 6971...

Page 8: ...in which case the user will be required to correct the interference at his own expense If you have a Class B digital device please read and follow the FCC information given below However there is no g...

Page 9: ...ts the balance scale must warm up for 30 minutes before operation Only after this time will the instrument have reached the required operating temperature Using Verified Balances Scales in Legal Metro...

Page 10: ...rectangular weighing pan Turn the 2 front feet as shown in the diagram until the air bubble is centered within the circle of the level indicator In most cases this will require several adjustment step...

Page 11: ...g is not permitted in legal metrology Open cover plate on the bottom of the balance scale Important set the balance scale on its side to access the cover plate DO NOT turn the balance scale upside dow...

Page 12: ...mbols for active application W Z L V R A C Position Designation 13 Delete Clear Function This key is generally used to cancel functions Quit application program Cancel calibration adjustment routine E...

Page 13: ...te and adjust it at the place of use using the built in motorized calibration weight for details see Calibration Adjustment in this chapter The temperature range C indicated on the verification label...

Page 14: ...the balance scale e 0 0 g Self test is performed followed by automatic initial tare function 2 Place container on weighing pan 11 5 g in this example 11 5 g 3 Tare the balance scale w 0 0 g 4 Place sa...

Page 15: ...alibration adjustment Cal Unit Grams Kilogr or Pounds not for verified models External Calibration in Verified Balances Scales When the balance scale is used in legal metrology external calibration is...

Page 16: ...CE suffix Set the following parameters Setup Bal Scal Cal Adj Cal Int menu code 1 1 9 4 The built in motorized calibration weight is applied and removed automatically for inter nal calibration Step K...

Page 17: ...Q Cal Ext Once you store the zero point 5000 0 g the required calibration weight is prompted flashing display 3 Apply the prompted calibration 5000 0 g weight in this example 5000 g Weight too light a...

Page 18: ...requirements Functions of the Keys during Configuration Symbol Key Function V b Scroll through menu items u One menu level lower u Confirm menu item c press and hold Save settings and exit menu from...

Page 19: ...languag code is displayed again 3 Select menu level u English o scrolls to the right 4 Change setting b Espanol Scroll until the desired setting is shown 5 Confirm the menu code u Espanol o o indicate...

Page 20: ...Additional functions Signal Acoustic signal beep 1 8 2 Keys Keypad 1 8 3 Ext key External switch function 1 8 4 On Mode Power on mode 1 8 5 Backlit Display backlighting 1 8 6 Reset Menu Factory settin...

Page 21: ...t 1 1 3 5 8 Digit1 1 1 3 6 Taring 1 W o Stb W o stability 1 1 5 1 Taring W StAb After stability 1 1 5 2 AutoZer Off 1 1 6 1 Auto zero On 1 1 6 2 Wt Unit For list of units see 1 1 7 1 Basic weight Togg...

Page 22: ...W stability 1 6 1 2 automatic Aut W o Autom w o stability 1 6 1 3 Aut With Autom w stability 1 6 1 4 StopAut Stop Off Not possible 1 6 2 1 automatic printing On Use print key r 1 6 2 2 Aut cycl EachVa...

Page 23: ...6 8 2 date dd mmm yy Day month year 1 6 9 1 mmm dd yy Month day year 1 6 9 2 Extras Menu CanEdit Can change settings 1 8 1 1 Additional Rd Only Read only 1 8 1 2 functions Signal Off 1 8 2 1 Acoustic...

Page 24: ...intout Total2 Comp Prt Off 2 6 1 1 Totalizing Component On 2 6 1 2 printout AnimalW 2 Activty Calm Fluct 2 of test obj 2 7 1 1 Animal Animal activity Active fluct 5 of test obj 2 7 1 2 weighing V Acti...

Page 25: ...ion of the Keys when Entering ID Numbers b key Press and hold to repeat Display Key Display Function symbol First position u Go to next position b V Select current position c Exit without saving chang...

Page 26: ...odes Languag English factory setting 5 1 Deutsch German 5 2 Franc French 5 3 Ital Italian 5 4 Espanol Spanish 5 5 pycckv v Russian 5 6 polski Polish 5 7 Codes Menu shows codes not texts 5 8 Applicatio...

Page 27: ...quantity is stored in battery backed memory Reference Sample Updating Automatic reference sample updating optimizes the counting accuracy You can activate or deactivate this function in the menu Auto...

Page 28: ...ents of 1 1 2 3 etc to 100 Press briefly In increments of 10 10 20 etc to 100 b press ref 20pcs and hold 6 Confirm selected reference sample u 20pcs quantity and start application nRef 20 pcs The curr...

Page 29: ...e percentage is stored in battery backed memory Preparation Select the Weighing in percent application in the menu see Configuration Set the following parameters Applic Application program Percent Wei...

Page 30: ...ance scale in this example 111 6 g 4 Initialize the balance scale u 100 0 The current reference weight pRef 100 remains stored until a new reference Wxx 111 6 g is set or the power supply is interrupt...

Page 31: ...f 1 Press the b key briefly To increase the value without pressing repeatedly Press and hold the b key The selected operator is stored in battery backed memory Preparation Select the Calculation appli...

Page 32: ..._____0 3 Set the divisor in this example 0 06237 Position the decimal point u 5x b __ 00000 2 u Enter numerals Repeatedly __ 06000 or press and hold b u etc __ 06237 4 Store the divisor and initialize...

Page 33: ...Select the desired number of measurement 1 to 100 In increments of 1 Press the b key briefly In increments of 10 Press and hold the b key The selected number of measurements is stored in battery back...

Page 34: ...to 100 b press ref 20 and hold 5 Confirm number of measurements u 0 0 g and start automatic animal weighing The number of measurements remains stored in battery backed memory until the setting is chan...

Page 35: ...ar component memory following cancellation of the weighing sequence by pressing c and printout of the total weight Toggling between component weight and total weight by pressing and holding b 2 sec Pr...

Page 36: ...nce scale w 0 0 g 3 Add first component 120 5 g 4 Store component data u 0 0 g NET Comp 1 120 5 g 5 Add next component 70 5 g NET 6 Store component data u 0 0 g NET Comp 2 70 5 g 7 Weigh in further co...

Page 37: ...rrent individual weight value and the value in totalizing memory by pressing b Printout of the total of the individual component weights S Comp To close the application program and print the total wei...

Page 38: ...his example 380 g 3 Store value in memory u 380 0 g Comp 1 380 0 g 4 Remove sample 0 0 g 5 Place the next sample on the balance scale in this example 575 g 575 0 g 6 Store value in memory u 955 0 g 57...

Page 39: ...lb and then to Troy ounces ozt Set the following parameters Applic Unit code 2 2 Step Key or instruction Display Data output Preparation 1 Begin selection of an application weight unit b None o 2 Sele...

Page 40: ...45544638 tl 10 Twn Tael Taiwanese taels 0 02666666000 tl 11 Grains Grains 15 4323583500 GN 12 Peny Wt Pennyweights 0 64301493100 dwt 13 Milligr Milligrams 1000 00000000 mg 14 Pt P Lb Parts per pound 1...

Page 41: ...wise an error message is displayed You can have results displayed with one decimal places or no decimal places see Configuration Note the sample holder and suspension wire used in the example below ar...

Page 42: ...8 0 99866 0 99864 18 0 99862 0 99860 0 99859 0 99857 0 99855 0 99853 0 99851 0 99849 0 99847 0 99845 19 0 99843 0 99841 0 99839 0 99837 0 99835 0 99833 0 99831 0 99829 0 99827 0 99825 20 0 99823 0 998...

Page 43: ...briefly or press and hold u etc _0 99823 5 Save density value and start application u The density value is stored in battery backed memory 6 Confirm Air display u Air 7 Determine the weight of the sam...

Page 44: ...ent Setup Prnt Out GLP Cal Adj menu code 1 6 7 2 or ISO GLP compliant printout or record always on Setup Prnt Out GLP Always on code 1 6 7 3 Line format for printout include data ID codes 22 character...

Page 45: ...ld for operator signature Blank line Dotted line ISO GLP compliant printout for external calibration adjustment Dotted line 17 Aug 2006 10 30 Date time beginning of measurement SARTORIUS Balance scale...

Page 46: ...iversal remote switch Pin 16 Not connected Pin 17 Not connected Pin 18 Not connected Pin 19 Not connected Pin 20 Data Terminal Ready DTR Pin 21 Not connected Pin 22 Not connected Pin 23 Not connected...

Page 47: ...e Center err 02 Calibration parameter Calibrate only when zero not met e g is displayed balance scale not tared Press w to tare the balance scale load on weighing pan Unload the balance scale err 10 T...

Page 48: ...aning agents on the following parts keypad overlay power jack data interface or any other plastic parts After cleaning wipe down the balance scale with a soft dry cloth On analytical balances remove a...

Page 49: ...the aim of recycling it In Germany and many other countries Sartorius takes care of the return and legal ly compliant disposal of its electrical and electronic equipment on its own These products may...

Page 50: ...48 60 Hz Power consumption VA maximum 16 typical 8 STNG6 Approx hours of operation with the YRB05Z rechargeable battery pack backlighting on h 35 Ambient Conditions The specifications given here are...

Page 51: ...size mm 90 d 90 d 90 d Whg chamber height mm 230 230 160 Dimensions W D H mm 230 303 330 230 303 260 Model GK1203 GK703 GK303 GK703 ST Weighing capacity 1200 ct 700 ct 300 ct Readability 0 001 ct 0 00...

Page 52: ...3 2 3 6 3 2 3 6 4 4 3 2 3 6 4 4 Weighing pan size mm 115 d 115 d 115 d Dimensions W D H mm 230 303 230 303 136 136 ED DS 230 303 330 Model ED153 GK3102 GK2202 ED153 CW ED153 DS Weighing capacity 150 g...

Page 53: ...F1 500 F2 Net weight approx kg 33 1 3 1 3 5 3 1 3 5 3 1 3 1 3 5 3 1 3 5 2 2 6 Weighing pan size mm 180 180 180 180 180 180 180 180 150 d Dimensions W D H mm 230 303 91 230 303 87 Model ED8201 GW7201 E...

Page 54: ...size mm 90 d 90 d 90 d 90 d Weighing pan size mm 230 230 160 160 Dimensions W D H mm 230 303 330 230 303 260 Model ED623S 0CE ED523S PCE ED423S 0CE ED323S 0CE ED153 0CE ED153 PCE Type BD ED 200 BD ED...

Page 55: ...ure Net weight approx kg 3 5 GW 0CEDS 4 9 Weighing pan size mm 180 180 GW 0CEDS 150 d Dimensions W D H mm 230 303 91 GW 0CEDS 230 303 260 Model ED822 0CE ED8201 0CE ED5201 0CE ED2201 0CE Type BD ED 20...

Page 56: ...ee Specifications for hours of operation To recharge the battery pack Unplug the AC adapter from the balance scale and plug it into the battery pack SartoConnect 1 data transfer software for direct tr...

Page 57: ...amples and containers 100 V to 230 V 50 60 Hz YSTP01 Anti vibration balance scale table for precise reliable weighing operations YWT01 made of cast stone with shock absorbers YWT03 Bracket for wall mo...

Page 58: ...58...

Page 59: ...59...

Page 60: ...60...

Page 61: ...61...

Page 62: ...62...

Page 63: ...ttingen Germany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius The status of the informati...