80 Cubis
®
MCA Operating Instructions
Technical Data
15.8 Recommended Calibration Weight
MCA14202S MCA14202P
MCA10202S
MCA8202S
Unit
Value
Value
Value
Value
External test weight
g
14000
14000
10000
7000
Recommended accuracy class
E2
E2
E2
E2
MCA6202S
MCA6202P
MCA4202S
MCA2202S
Unit
Value
Value
Value
Value
External test weight
g
5000
5000
3000
1500
Recommended accuracy class
E2
E2
E2
E2
MCA1202S
MCA12201S
MCA8201S
MCA5201S
Unit
Value
Value
Value
Value
External test weight
g
700
12000
8000
5000
Recommended accuracy class
E2
F1
F1
F1
15.9 isoCAL Function
15.9.1 Models MCA14202S | MCA14202P | MCA10202S
Unit
Value
isoCAL is triggered by the following criteria:
In the event of a temperature change
K
1 .5
After a time interval
h
12
After successful leveling
15.9.2 Models MCA8202S | MCA6202S | MCA6202P | MCA4202S | MCA2202S | MCA1202S
Unit
Value
isoCAL is triggered by the following criteria:
In the event of a temperature change
K
2
After a time interval
h
12
After successful leveling
15.9.3 Models MCA12201S | MCA8201S | MCA5201S
Unit
Value
isoCAL is triggered by the following criteria:
In the event of a temperature change
K
4
After a time interval
h
12
After successful leveling