Cubis
®
MCA Operating Instructions
63
Operation
8.10 Internally Calibrating and Adjusting the Device
Requirements
The weighing pan is unloaded .
Procedure
t
Open the main menu .
t
Press the “Internal Adjustment” task .
y
The internal calibration / adjustment function is executed .
y
If automatic leveling is set for a model with motorized adjusting feet: The device
levels itself automatically .
y
If the calibration / adjustment function is complete: The device confirms the
completion of the calibration / adjustment process with an acoustic signal, and
the calibration report is displayed .
t
To close the calibration report and return to the main menu: Press the [OK]
button .
8.11 Weighing and Printing with ID Marking
8.11.1 Saving Values for Printouts
Every lot and every sample can be assigned an ID number . The ID numbers are saved
in the print memory and exported during the printing process .
Requirements
The prompt for the lot IDs and sample IDs is activated for the “Weighing” task .
Procedure
t
Open the main menu .
t
Start the “Weighing” task: To do so, tap on the task .
y
The input field for the lot ID is displayed .
t
If a barcode scanner is connected to the device: Enter the lot ID in the entry field
or scan using the barcode scanner .
t
If
no
barcode scanner is connected to the device: Type the lot ID into the entry
field .
t
Press the [OK] button .
y
The weighing display is displayed .
t
Zero the device . In order to do so, press the [Zero] button .
t
Place the sample on the weighing pan .
t
Press the [Confirm] button .
y
The input field for the sample ID is displayed .
t
Type the sample ID into the entry field .
t
Press the [OK] button .
y
The [Print memory] button appears in the operating display .
y
The weight value and entered IDs are saved .
t
If additional values are to be saved:
t
Remove the sample being weighed .
t
Place the next sample on the weighing pan and press the [Confirm] button .
t
Type the sample ID into the entry field .
t
Press the [OK] button .
Summary of Contents for Cubis MCA10202S Series
Page 2: ......
Page 85: ...Cubis MCA Operating Instructions 85 Conformity Documents...