Cubis
®
MCA Operating Instructions
59
Operation
8.3 Leveling the Device
8.3.1 Level Device with Motorized Leveling Feet
Leveling compensates any inclines at the device’s installation site . If leveling is
necessary: The [Leveling] button appears in the weighing display and a message
appears in the Status Center .
Procedure
t
If the weighing display is displayed: Press the [Leveling] button .
t
When the Status Center is displayed: Press the [Level] button .
y
The Leveling Wizard opens .
t
Follow the wizard’s instructions .
8.4 Activating Applications and Adding a Task
8.4.1 Activating Applications
All applications from the QAPP package “Essentials” are activated for the device at
the factory . Additional applications may be activated in the QAPP center .
Procedure
t
Open Task Management .
t
Press the [QAPP center] button .
y
An overview of all available QAPP packages is displayed .
t
Select the desired software application package e .g . pharmaceuticals .
t
To scroll through the contents of a displayed QAPP package: Swipe the contents
of the QAPP package upwards or downwards .
t
Select the desired QAPP package:
y
A list of all the applications contained in the QAPP package is displayed .
t
If the selected QAPP package is to be approved with all the applications it
contains or all QAPP packages are to be approved:
t
Press the [License] button .
y
The input field for the license key appears .
t
If an additional cost is associated with the QAPP package: Enter the license
key in the entry field and press the [OK] button .
t
If no additional cost is associated with the QAPP package: Press the [OK]
button .