Cubis
®
MCA Operating Instructions
53
System Settings
7.5.10 Parameters in the “Connections / Interfaces / COM-RS232 Interface” Menu
Parameter
Settings
Explanation
Protocol
Off*
Deactivates the RS232 connection .
SBI
Enables SBI communication . The data is output to a PC or control unit .
Enables the use of ESC commands from a PC to control the basic balance
functions with ASCII protocol .
xBPI
Extended range of commands to control numerous balance functions with
binary protocol for direct communication with the weighing module .
Log communication
Off*
Deactivates the automatic data recording for this interface .
On
Activates the automatic data recording for this interface .
Baud rate
600 baud
Sets the baud rate to 600 baud .
1200 baud
Sets the baud rate to 1200 baud .
2400 baud
Sets the baud rate to 2400 baud .
4800 baud
Sets the baud rate to 4800 baud .
9600 baud*
Sets the baud rate to 9600 baud .
19200 baud
Sets the baud rate to 19200 baud .
38400 baud
Sets the baud rate to 38400 baud .
57600 baud
Sets the baud rate to 57600 baud .
115200 baud
Sets the baud rate to 115200 baud .
Data bits
7 data bits
Sets the number of data bits to 7 .
8 data bits*
Sets the number of data bits to 8 .
Parity
Odd*
Applies an odd parity .
Even
Applies an even parity .
None
Does
not
apply a parity .
Stopbits
1 stop bit*
Sets the number of stop bits to 1 .
2 stop bit
Sets the number of stop bits to 2 .
Handshake
Software
Sets the handshake protocol to software handshake .
Hardware*
Sets the handshake protocol to hardware handshake .
None
Does
not
set a handshake protocol .
* Factory setting
7.5.11 Parameters in the “Connections / Interfaces / USB-B Interface” Menu
Parameter
Settings
Explanation
Protocol
Off*
Deactivates the USB-B connection .
SBI
Enables SBI communication . The data is output to a PC or control unit .
Enables the use of ESC commands from a PC to control the basic balance
functions with ASCII protocol .
SICS
Enables SICS communication . The data is output to a PC or control unit .
Enables the use of commands from a PC to control the basic balance
functions with SICS protocol .
xBPI
Extended range of commands to control numerous balance functions with
binary protocol for direct communication with the weighing module .
PC direct
Enables data output to a spreadsheet program via a direct PC connection .
Log communication
Off*
Deactivates the automatic data recording for this interface .
On
Activates the automatic data recording for this interface .
* Factory setting