12
Start up of cassettes
5.2 Clamping Forces
Sartocon
®
Slice 200 and Sartocon
®
Slice Holding Device, Manual closure
Cassette type
Order
number
Previous Slice
holder
and Slice 200
holder
New Slice 200
and Slice
holder with
trapezoidal
thread,
PEEK washer
and bronze nut
(available since
July 2014**)
Sterilization*
Torque applied
with a torque
wrench
Torque
applied
with a
torque
wrench
Torque
applied
with a
torque
wrench
Sartocon
®
Slice 200 and
Sartocon
®
Slice
Hydrosart
®
UF
308144…
305186…
20 Nm
17 Nm
t.b.d.
Sartocon
®
Slice 200 and
Sartocon
®
Slice
PESU UF
308146…
305154...
25 Nm
17 Nm
t.b.d.
Sartocon
®
Slice 200 and
Sartocon
®
Slice
PESU Max
308146AL…
305146AL…
25 Nm
25 Nm
No thermal
sterilization
Sartocon
®
Slice 200 and
Sartocon
®
Slice
Hydrosart
®
MF
308186…
305144…
20 Nm
25 Nm
t.b.d.
Sartocon
®
Slice 200 and
Sartocon
®
Slice
PESU MF
308154…
305146…
25 Nm
17 Nm
t.b.d.
* List of sterilizable cassettes is available in “Technical Specifications”, section 6.1.
Sartocon
®
2 plus Holding Device, Manual closure
Cassette type
Order
number
Required
Compression
Sterilization* Torque
applied
on Tie Rods
with a
Torque
Wrench
Sterilization*
Operating Conditions
Operating Conditions
Sartocube
®
and
Sartocon
®
Cassettes
302146... |
302154…
18–35 kN
< 10 kN
80 Nm
first notch
PESUmax
302146AL… 21–35 kN
No thermal
sterilization
80 Nm
No thermal
sterilization
Sartocube
®
and
Sartocon
®
Hydrosart
®
MF | UF
302144… |
302186…
18–25 kN
< 10 kN
60 Nm
first notch