Sanwa YX360TRF Instruction Manual Download Page 1

Black

D-1034-236 取扱説明書 YX360 TRF

オモテ・ウラ どん天

*1 Not including the resistance of fuse.
*2 The maximum value when the pointer was moved by
charged current in the capacitor.

DCV

CV

DCA

CV

0.1

±5% against
full scale

±5% against
full scale

±4% against
full scale

±3% against
full scale

±3% against
full scale
±3% against
full scale

±3% of arc

±5% of arc

Approximate Value

Approximate Value

Approximate Value

Input impedance 20k

1

/V

Input impedance 40k

1

/V

Input impedance 9k

1

/V

Current across test pins

Input impedance 9k

1

/V

30Hz ~ 100kHz within
±3% f.s. (AC10V range)

Center value 20

1

Max. value 2k

1

Release voltage 3V

*1 Voltage drop 0.1V

*1 Voltage drop 0.25V

*2

Input impedance 9k

1

/V

250/1000

10/50/250/750

50μ

2.5m/25m/0.25

2k/20k/200k/2M
(X1/X10/X100/X1K)

-10dB ~ +22dB
(for 10VAC)
~ +62dB

0 ~ 150mA at X1 range
0 ~ 15mA at X10 range
0 ~ 1.5mA at X100 range
0 ~ 150 μA at X1K range
0 ~ 1.5 μA at X100K range

200M
(X100K)
10μF

DC25kV

1000 atX10 range

HV-10T probe

HFE-6T probe

±5/±25

DCV
(NULL)

ACV ~

C

dB

LI

HV (DC 
high volt)

h

FE

Use the optional probe

1

YX360TRF

MULTITESTER

INSTRUCTION MANUAL

02-0807 2040 2040

SANWA ELECTRIC

INSTRUMENT CO.,LTD.

Dempa Bldg., 4-4 Sotokanda 2-Chome

Chiyoda-Ku,Tokyo,Japan

NAMES OF COMPONENTS

− 1 −

− 2  −

− 3  −

INTRODUCTION

Thank you for purchasing a SANWA tester Model YX360TRF. You are
kindly requested to thoroughly read this manual before use for safety.
Especially, “SAFETY INFORMATION” and “MEASURING PROCE-
DURE” are important. Keep this manual together with the tester so as
no to lose it.

Fig. 1

SAFETY INFORMATION

SPECIFICATIONS

The following are precautions to prevent accidents such as electrical
shocks.
Be sure to read them before using the tester.

Symbols

The following cautionary signs appear on the multitester and
in this manual.

Disobedience to instructions with this sign may lead to trouble
with the tester and accidents such as electrical shock.

This sign cautions that high voltage is applied to parts marked
with it.

Precautions for Safety Measurement

To ensure that the meter is used safely, follow all safety and operat-

ing instructions.

1

1. Never use the meter on the electric circuits that exceeds 3kVA.

1

2. Pay special attention when measuring the voltage of AC 33 Vrms

(46.7V peak) or DC 70V or more to avoid injury.

1

3. Never apply input signals exceeding the maximum rating input

value.

1

4. Never use the meter for measuring the line connected with equip-

ment (i.e. motors) that generates induced or surge voltage since it
may exceed the maximum allowable voltage.

1

5. Never use the meter if the meter or test leads are damaged or broken.

1

6. Never use an uncased meter.

1

7. Be sure to use a fuse of the specified rating or type. Never use a

substitute of the fuse or make a short circuit of the fuse.

1

8. Always keep your fingers behind the finger guards on the probe

when making measurements.

1

9. Be sure to disconnect the test pins from the circuit when changing

the function or range.

10. Before starting measurement, make sure that the function and

range are properly set in accordance with the measurement.

11. Never use the meter with wet hands or in a damp environment.
12. Never use test leads other than the specified type.
13. Never open the case except when replacing batteries or fuses. Do

not attempt any alteration of original specifications.

14. To ensure safety and maintain accuracy, calibrate and check the

meter at least once a year.

15. Indoor use only.

WARNING

− 5  −

− 6  −

− 7  −

BODY COVER, TEST LEADS, HAND STRAP

Use of Cover 

(example for the body cover)

When this tester is out of use:
Attach the cover to the panel face for safekeeping.

When measuring:
Attach it either to the rear case side
or use it as a stand as shown below.

Storage of Test Leads

When placing the test leads in the
storage space, roll it 3 times, then
put in the test pin side first as shown
below.

Attachment of Hand Strap

1 Loosen the screws fixing the rear

case and remove it.

2 Hand strap is attached to con-

necting point.

3 Put back the rear case where it

was and fix it with the screws.

Fig. 2

− 4  −

SCALE READING

Range

Multiplied

1

X 100k

X 100k

1

X 1k

X 1k

1

X 100

X 100

1

X 10

X 10

1

X 1

X 1

DCV 250

X 1

DCV 2.5

X 0.01

DCV 0.25

X 0.001

ACV 250

X 1

DCA 0.25

X 0.001

DCA 25m

X 0.1

DCA 2.5m

X 0.01

DCV 50

X 1

ACV 50

X 1

DCA 50 μ

X 1

DCV 0.1

X 0.01

Range

Multiplied

DCV 10

X 1

DCV 1000

X 100

ACV 750

X 100

ACV10

X 1

C (μF)

X 1

DCV ± 25

X 1

DCV ± 5

X 1

150mA at X 1

X 10 

15mA at X 10

X 1

1.5mA at X 100

X 0.1

150 μA at X 1k

X 10

1.5 μA at X 100k

X 0.1

LV

X 1

h

FE

X 1

ACV 10

X 1

ACV 50

14dB added

ACV 250

28dB added

ACV 750

40dB added




  


General Specifications

(Temperature : 23±2˚C humidity 75% RH max. No condensation)

Drop shock proof

A taut-band structure is adopted in the meter part.
The meter part is designed to withstand shock.

Circuit protection

The circuit is protected by fuse even when voltage of up
to AC 230V is impressed on each range for 5 seconds.

Internal battery

Internal fuse

R6 (IEC) or UM-3  1.5V X 2

F500mAH/250V  Ø5.2 X 20mm  Fast acting fuse

Operating temperature
and humidity range

5 ~ 31˚C, 80%RH max.
31< ~ 40˚C, 80 ~ 50%RH (decreasing linearly)

Withstand voltage

Dimensions and weight

6kV AC (1min.) between input terminal and case

159.5 X 129 X 41.5 mm/ approx. 320g

Storage temperature/
Humidity

-10 ~ 50˚C 70%RH max. No condensation

Accessories

Instruction manual 1, Hand strap 1,

Items

Specification

Note

: The definition of installation category,i.e.

CAT II : Local level, appliances, portable equipment etc.,
with smaller transient overvoltages than installation cate-
gory III.
CAT III : Distribution level, fixed installation, with smaller
transient overvoltages than installation category IV.

Application

This instrument is portable multitester designated for meas-
urement of weak current circuits.

The specifications described in this manual are subject to
change without notice.

Measurement Range and Accuracy

Function

Full scale value

Accuracy

Remarks

0.25/2.5/10/50

±3% against
full scale

APPLICATION

• HV probe, HV-10T  • h

FE

probe, HFE-6T  • Test lead for repair, TL-6IT

8  −

9  −

10  −

11  −

MEASURING PROCEDURE

Measuring DCV

1

Set the range selector knob to

an appropriate DCV range.

2

Apply the black test pin to the

negative potential, and the red

test pin to the positive poten-

tial of the circuit.

3

Read the movement of the

pointer by V and A scale.

WARNING

Confirm the range of use before measurement.

Preparation for Measurement

1

Adjustment of meter zero position:

Turn the zero position adjuster so that the pointer may align

right to the zero position.

2

Range selection:

Select a proper range for the item to be measured and set

the range selector knob accordingly.

When determining a measuring range, select a one higher

voltage than the value to be measured as well as where the

pointer of a meter moves to a considerable extent.

However, select the maximum range and measure in case

the extent of value to be measured can not be predicted.

NOTE

Measuring ±DCV (NULL)

1

Set the range selector knob to

an appropriate ±DCV (NULL)

range.

2

Turn the 0

1

adjuster so

th

at

 t

he

 p

oi

nt

er

 m

ay

 a

li

gn

exactly to 0 by ±DCV scale.

3

Apply the black test pin to the

negative potential side, and

the red test pin to the positive

potential side of the circuit.

4

Read the movement of the pointer by ±DCV scale.

Fig. 3

Measuring ACV 

1

Turn the range selector knob to an appropriate ACV range.

2

Apply the test leads to the circuit to be measured.

3

Read the movement of the pointer by V and A scale. (Use

AC 10V scale for 10V range only.)

•Since this instrument provides

the mean value system for its AC

voltage measurement circuit, AC

waveform other than sine wave

may cause an error.

•Errors occur under such frequen-

cies other than those in the speci-

fication table.

Measuring DCA

1

Turn the range selector knob to an appropriate DCA range.

2

Take out the circuit to be measured and apply the black test

pin to the negative potential,

and the red test pin to the

positive potential of the cir-

cuit.

3

Read the movement of the

pointer by V and A scale. 

WARNING

Connect the meter in series with the load.

Cut off.

Fig. 4

Fig. 5

Measuring 

1

1

Turn the range selector knob to an appropriate 

1

range.

2

Short the red and black test pins and turn the 0

1

adjuster so

that the pointer may align exactly to 0

1

. (If the pointer fails

to swing up to 0

1

even when the 0

1

adjuster is turned

clockwise fully, replace the internal battery with a fresh one.)

3

Apply the test pin to measur

resistance.

4

Read the movement of the point-

er by 

1

scale.

Note

:The polarity of + and - is

reverse to that of the test

leads when measurement is

done in 

1

range.

Note

:How to replace battery.

1

Loosen the screws fixing the rear

case and remove it.

2

Replace R6 (UM-3) to fresh dry

batterise.

3

Put back the rear case where it

was, and fix it with the screws.

Note

:Be sure to use the same rated

fuse. In case a fuse other than

the same rate (see “SPECIFICATIONS“) is used, an error in

indication occurs and/or circuit protection is made unable.

WARNING

Do not measure the resistance in a circuit where

a voltage is present.

Fig. 6

Fig. 7

Measuring Capacity (C)        

1

Set the range selector knob to C(μF) .

2

Measure the capacitance by applying

the test pin to the capacitor to be meas-

ured after the 0

1

adjustment is made

in the same manner as in the resist-

ance measurement.

3

The pointer moves full scale by the charge current to the

capacitor. However, the pointer gradually starts returning to its

original position. Read the indicated maximum value on C(μF)

scale.

Note

:Be sure to short circuit the both ends of the capacitor for

discharge prior to the initial measurement or in such case

after the measurement is made.

Note

:Pay due attention to the polarity (+ and -) of the capacitor.

(C side of the capacitor to - side of the meter.)

Measuring AF Output (dB)        

dB (decibel) is measured in the same way as the ACV

measurement, but by reading the dB scale instead.

For measurement on the 10V range, the dB scale (- 10dB ~

+22dB) is read directly, but, when measured on the 50V

range, 14dB is added. On the 250V range, 28dB is added to

the reading on the scale, and on the 1000V range, 40dB

added.

Thus, the maximum dB readable is 22 + 40 = 62 (dB) meas-

ured on the 1000V range.

Note

:Cut direct current with a capacitor of 0.1μF or more when

measuring such signal as having direct current.

Fig. 8

12  −

13  −

14  −

15  −

MEASURING PROCEDURE

Measuring of Iceo (Leak Current) for Transistor

1

Adjust 0

1

by setting the range selector

knob to a proper range from X1 ~ X1k.

2

For NPN transistor, apply a black test

lead to the collector and a red one to

the emitter.

For PNP transistor, the red one to the col-

lector and the black one to the emitter.

3

Determine the leak current by ICEO

scale indicated on the scale plate. (Unit

in

 μA, mA)

Measuring of Diode (including LED)

1

Adjust 0

1

by setting the range selector knob to a proper

range from X1 (150mA) ~ X100k (1.5μA).

2

Apply the black test lead to anode side and the red one to

cathode side when measuring IF (forward current). Apply

the black test lead to cathode side and the red one to anode

side when measuring IR (reverse current).

3

Read the indicated value by LI

scale. (The pointer moves to a

considerable extent for IF, and

little extent for IR)

4

The value indicated on LV scale

during the measurement is the

forward voltage of diode.

NPN

B

C

bla

ck

red

E

PNP

B

C

red

bla

ck

E

red

bla

ck

A                     K

I

F

I

A

Fig. 9

Fig. 10

USAGE OF OPTIONAL PROBES

Usage of High Voltage Probe (HV-10T)

Up to DC 25kV of CRT anode voltage can be measured by

connecting optional HV-10T probe.

WARNING

Keep the hand (finger) away from high voltage

power supply. Electric shock may occur due to dis-

charge.

Measurement should be limited only to micro cur-

rent circuits.

1

Turn the range selector knob and set it to HV PROBE (DC

2.5V range).

2

Connect the jack of the black lead of the probe to the black

test pin, and the jack of the red lead to the red test pin.

3

Apply the probe clip to the earth side and the measuring pin

to the point to be measured.

4

Read out measured value on 0~

 250 of V scale in kV unit

after multiplying it by 0.1.

Usage of h

FE 

PROBE (HFE-6T)

1

Set the range selector knob to X10 range ( h

FE  

PROBE).

2

Short circuit both the red and black test pins to adjust 0

1

.

3

Connect the black test pin to the probe jack when a transis-

tor to be measured is NPN, and the red pin to the probe

jack for PNP transistor.

4

Connect the black clip of the probe to the transistor base

and the red clip to the collector.

5

Connect the remaining test lead to the emitter and measure

h

FE

.

6

Read the indicated value of the meter on h

FE

scale.

MAINTENANCE

How to Replace the Fuse

If voltageover 100V is applied

to DCA and 

1

ranges, the fuse

may blow out to protect

th

e

circuit.

1

Loosen the screws fixing the

rear case and remove it.

2

Pull the fuse out of holder on the

circuit board and replace it.(Fig. 11)

3

Put back the rear case where it was and tighten the screws.

4

Check and see whether or not indications of respective

ranges are normal (check other parts for any failures).

Storage and Other Precautions

1

Avoid giving the meter any excessive shock or vibration by

loading it on the motorbike, for instance.

2

Keep off dust and moisture from the meter.

3

Do not leave the meter for a long time in places of a high

temperature (higher than 55˚C), a high humidity (higher than

80%), and dew condensation.

4

The meter cover is treated with antistatic coating. Do not

wipe it hard or clean it with volatile solvent. Use a soft

brush to remove dust.

Fig. 11

AFTER - SALES SERVICE

Warranty and Provision 

Under Sanwa's general warranty policy, each instrument is war-

ranted to be free from defects in workmanship or material

under normal use for the period of one (1) year from the date of

purchase.

This warranty policy is valid within the country of purchase only,

and applied only to the product purchased from Sanwa author-

ized agent or distributor.

This warranty does not apply to fuses, disposables batteries, or

any product or parts, which have been subject to one of the fol-

lowing causes:

1.A failure due to improper handling or use that deviates

the instruction manual.

2.A failure due to inadequate repair or modification by peo-

ple other than Sanwa service personnel.

3.

A failure due to causes not attributable to this product

such as fire, flood and other natural disaster.

4.Non-operation due to a discharged battery. 

5.A failure or damage due to transportation, relocation or

dropping after the purchase.

Repair

Please contact Sanwa authorized agent/distributor/service

provider, listed in our website, in your country with your

information.

SANWA web site

http://www.sanwa-meter.co.jp

E-mail: [email protected]