47/52
9.8 Second Hand Usage
All second-hand units MUST be inspected for a safety check and attendant certified
by an authorized specialist.
All operators must have proof of having been trained in how to operate the
stairclimber.
IMPORTANT:
In addition to the intervals, described above, all safety checks must be carried out in
full each time there is a change of operator and each time the stair-climber is put into
operation after a period of down time, even if the stair-climber appears to be
undamaged and operating normally at first sight.
10
Troubleshooting
Problem
: LED is red and stairclimber does not switch on
Cause
: The stairclimber is at too flat an angle and the level switch is preventing it
from switching on
Problem
: The main switch does not work even though the handle unit is fitted.
Cause
: The lever screw has not been tightened enough, or the handle unit is
mounted back-to-front.
Problem
: The climber unit does not run smoothly on the main wheels; it keeps on
braking!
Cause
: The lifting frame with the support wheels is not in its neutral position between
the main wheels. Move forwards or backwards using single-step mode (see section
4.1.3) until the lifting frame stops in the neutral position.