Advice on use
• In order to ensure comparable values, always measure
your blood pressure at the same time of day.
• Before every measurement, relax for about five minutes.
• If you want to perform several measurements on the same
person, wait five minutes between each measurement.
• Do not take a measurement within 30 minutes after eating,
drinking, smoking or exercising.
• Repeat the measurement if you are unsure of the measured
value.
• The measurements taken by you are for your information
only – they are not a substitute for a medical examination!
Discuss the measurements with your doctor, and never
base any medical decisions on them (e.g. medicines and
their administration)!
• Do not use the blood pressure monitor on newborns, preg-
nant women or patients with preeclampsia.
• In the case of restricted circulation on the arm as a result of
chronic or acute vascular diseases (including vascular con-
striction), the accuracy of the wrist measurement is limited.
In this case you should avoid using an upper arm blood
pressure monitor.
• Cardiovascular diseases may lead to incorrect measure-
ments or have a detrimental effect on measurement accu-
racy. The same also applies to very low blood pressure,
diabetes, circulatory disorders and arrhythmias as well as
chills or shaking.
• The blood pressure monitor must not be used in connec-
tion with a high-frequency surgical unit.
• Only use the unit on people who have the specified wrist
measurement for the device.
• Please note that when inflating, the functions of the limb in
question may be impaired.
• During the blood pressure measurement, blood circulation
must not be stopped for an unnecessarily long time. If the
device malfunctions, remove the cuff from the arm.
• Do not allow sustained pressure in the cuff or frequent
measurements. The resulting restriction of the blood flow
may cause injury.
• Ensure that the cuff is not placed on an arm in which the
arteries or veins are undergoing medical treatment, e.g.
intravascular access or therapy, or an arteriovenous (AV)
shunt.
• Do not use the cuff on people who have undergone a mas-
tectomy.
• Do not place the cuff over wounds as this may cause fur-
ther injury.
• The blood pressure monitor can only be operated with bat-
teries.
• To conserve the batteries, the monitor switches off auto-
matically if no buttons are pressed for one minute.
• The device is only intended for the purpose described in
these instructions for use. The manufacturer is not liable
for damage resulting from improper or careless use.
Storage and Care
• The blood pressure monitor is made up of precision elec-
tronic components. Accuracy of readings and the instru-
ment’s service life depend on careful handling.
13