EB 9510 EN
85
Annex
Menu
Adjustment range/values/description
Error message E1
4.2.2
Determines whether an error message is issued in case of
condensed state (E1) (see page 84).
▪
No
▪
Yes
Error message E2
4.2.3
Determines whether an error message is issued in case of
condensed state (E2) (see page 84).
▪
No
▪
Yes
Error message E3
4.2.4
Determines whether an error message is issued in case of
condensed state (E3) (see page 84).
▪
No
▪
Yes
Diagnostic data
4.3
Operation duration
4.3.1
Read the entire operating time of the device (dd:hh:mm:ss)
Temperature
4.4
Temperature inside device 4.4.1
Read the current device temperature in °C
Max. temperature inside
device
4.4.2
Set an upper temperature limit within the specified range. If
the current device temperature is above the adjusted limit, an
error message is generated and displayed. The status
changes to 'Out of specification'.
▪
10 to 80 °C
(default: 70 °C)
Min. temperature inside
device
4.4.3
Set a lower temperature limit within the specified range. If
the current device temperature is below the adjusted limit, an
error message is generated and displayed. The status
changes to 'Out of specification'.
▪
–40 to +10 °C
(default: –40 °C)
Filling level events
4.5
Filling limit alarm (SCN)
4.5.1
Activate/deactivate the permissible filling level
▪
ON
▪
OFF (default)
Pre-alarm
4.5.2
Activate/deactivate the pre-alarm when the filling level falls
below the limit.
▪
ON (default)
▪
OFF