background image

2

3

4

1

5

The

 

SAM

®

 XT

 

 

  

MUST NOT

 be applied 

over solid objects  

within the clothing.

  

CAUTION

Insert the tip of the strap through the 
opening in the buckle. Grip the strap 
close to the buckle and pull hard and 
steady against the buckle until it stops. 
Maintaining tension, press the strap 
back upon itself to hold it in place.

DO NOT COVER WINDLASS C-HOOKS

.

Twist the metal windlass rod until the 
bleeding has stopped.

Wrap strap between Windlass C-hooks, 
over rod and around the limb. 

Secure the ( ) Time Band over Windlass 
C-hooks and record time of application.

sammedical.com

Locate the site of injury

*

 

and place the

 

SAM

®

 XT

 

around the limb, directly onto 

the skin 2-3 inches above the wound.

If the site of the injury is not 
easily seen, place the 

SAM

®

   XT

 

as “high & tight” as possible on 
the injured arm or leg. When the 
situation permits, re-evaluate 
the injured limb and place the

 

SAM

®

   XT

 

directly onto the skin 

2-3 inches above the wound.

*

Fix the metal windlass rod inside the  
Windlass C-hooks and confirm that 
bleeding has stopped. If bleeding continues, 
try to further tighten the tourniquet with 
additional twists of the metal windlass rod. 
If this fails to stop the bleeding, apply a 
second 

SAM

®

 XT

 

side by side next to the 

first and ensure bleeding has stopped. 

SAM XT - EXTREMITY TOURNIQUET 

INSTRUCTIONS FOR USE

Reviews: