2.3 Menu Program Edit
The example assumes to edit the program #4 with 2 steps
Menu
_
Program
____
_
Edit
Edit an existing program
confirm menu choice
Ä
2.3.1
T
/
S
scroll through the Menu.
ESC
returns to
Ä
2
2.3.1 Choose program to edit
Program Step
P04 S01
+/-
select desired value
confirms the value and skips to the next field (P
Ä
S) or
Ä
2.3.2
(to the selected Step Number respective)
ESC
returns to
Ä
2.3
Note
: Only used program and step numbers will appear in the
display
2.3.2 Edit Temperature – Step 1
Set
_
Temperature
_
04/01
_____
90.0°C
+/-
change desired value. Pre-set value is the last used value in
manual mode.
confirms and saves the value
Ä
2.3.3
ESC
restores the old value or returns to
Ä
2.3
Note:
Display 04/01 means actual program/step number.
2.3.3 Edit Gradient – Step 1
Gradient
____
04/01
__
1.5°C/Min
+/-
change desired value
confirms and saves the value
Ä
2.3.4
ESC
restores the old value or returns to
Ä
2.3
2.3.4 Edit Holding Time – Step 1
Hold
_
Time
___
04/01
_
H:M
_
11:00
+/-
Change desired value
confirm value and skips to the next input field (HH
Ä
MM) or
stores the time and go to
Ä
2.3.5
ESC
restores the old value and skips back one input field
(M
Ä
H) or returns to
Ä
2.3
2.3.5 Edit Fan Speed – Step 1
Fan
_
Speed
___
04/01
_______
100%
+/-
change desired value
confirms and saves the value
Ä
2.3.6 (next step)
ESC
restores the old value or returns to
Ä
2.3
2.3.6 Edit Temperature – Step 2
Set
_
Temperature
_
04/02
____
110.0°C
+/-
change desired value. Pre-set value is the last used value in
manual mode.
confirms and saves the value
Ä
2.3.7
ESC
restores the old value or returns to
Ä
2.3
Note:
Display 04/01 means actual program/step number.