- 14 -
Copyright © SalvisLab / Edition 191107
1 Main Menu Temp & Options
Main Menu
Temp & Options
Manual operation with a set temperature. You can select options
like gradient, hold-time (dwell-time), fan speed, pre-setted start
date/time.
Press
to select
Ä
1.1
ESC
returns to
Ä
1
1.1 Set Temperature
Set
_
Temperature
_
_________
120.0°C
+/-
change desired value.
confirms and saves value
Ä
1.2
ESC
restores the old value or
returns to
Ä
1
1.2 Select Quick Start or Start with Options
Start
_
?
___
Now
__
Options
+/- Select desired answer
confirms and saves value
If
Now
selected: System will start immediately
Ä
6
If
Option
selected
Ä
1.3
ESC
cancels and
returns to
Ä
1
1.3 Set Gradient
Gradient
2,0°C/Min
+/-
Change value
confirms and saves the value
Ä
1.4
ESC
restores the old value or returns to
Ä
1
Note:
A value of
0,0
means maximal possible heating rate!
1.4 Set Holding Time (dwell time)
Hold Time
_______
HHH:MM
____
15:00
+/-
Change value
confirm value and skips to the next input field (HH
Ä
MM) or
stores the time and go to
Ä
1.5
ESC
restores the old value and skips back one input field
(MM
Ä
HH) or goes back to
Ä
1
Note
: A value of 0:00 means endless holding time
1.5 Set Fan Speed
Fan
_
Speed
____________
100%
+/-
Change value
confirms and saves the value
Ä
1.6
ESC
restores the old value or returns to
Ä
1
Note:
The minimal Fan Speed is depending on the system and is
factory set.