Fig.32 <Disp More> Secondary Submenu
<Range>
Setting display range
<Range> is used to set the largest scale of <Bar>, <Chart> and <Polar> chart.
The input range of <Range> is 1 nV to 1 V. This range will change in sync with the change of
<Sensitivity>. Besides, it can be set numerically through the keyboard. For example, 1.00e-01
means 100mV.
Keep <Sensitivity> unchanged and then change the value of <Range>. Set <Range>
1.00e-03=1mV, then the display range is -1mV to +1mV.
<Curve Point>
Setting the position of sampling curve.
In <Curve Point> mode, the <Cursor> in <Chart> can rapidly locate position and watch the value
of this point. The setting range of <Curve Point> is 1-16384.
When <Chart> is in reset mode (no store data), the <Curve Point> makes no sense and keeps 0.
When sampling data, this option will be hidden (can’t use). It will be displayed after pause.
Before using <Curve Point>, users should callout <Cursor> by pressing the [CURSOR] button in
CONTROL area.
3.5
[SAVE/RECALL]
Press [SAVE/RECALL] button and enter the submenu as is shown in Fig.33.