2. Set the parameters in the following sequence:
<Ref. Source>
Set to <Internal>
<Ref. Frequency>
Set to 997 Hz
<Sensitivity>
Set to <100 nV> through the spinning knob
<Reserve>
Set to <Low>
<Filter>
Set to <24 dB/oct>
<Trace>
Set <Trace> in <Display> to <Noise>
3. Wait until the reading is steady, and then record the maximum value of noise.
4. End the test and fill the result in the performance test form.
6.10
SE1022D Performance Test Record
SE1022D Performance Test Record
Serial Number
Tested By
_
Firmware Revision
Date
_
Equipment Used
_
_
1. Starting Tests
Pass
Fail
2. DC Offset
Input Coupling
Reading
Upper Limit
Channel A
AC
0.500mV
Channel A
DC
0.500mV
Channel B
AC
0.500mV
Channel B
DC
0.500mV
3. Common Mode Rejection
Frequency
Reading
Upper Limit
Channel A
100Hz
30μV
Channel B
100Hz
30μV